Instruction Manual
Bualo Filter Turbo Auto-Sense™ Operator’s Manual: Page 13
2.2 ViroSafe® Filter Instructions
Note: Before installing or removing any lter, be sure that the system is turned o.
Filter Installation Instructions:
e installation of the ViroSafe® Filter into the BUFFALO FILTER Turbo Auto-Sense™ Smoke
Evacuation System(s) is quick and simple.
1. Remove the ViroSafe® Filter from the shipping box and discard any protective wrapping. Ex-
amine all lters for damage during shipping and storage. Do not install any lter with visible
signs of structural damage.
2. Insert the ViroSafe® Filter into the lter receptacle. Be sure that the lter is seated completely
against the bottom of the lter chamber.
WARNING: is device is not intended for evacuation of uid. If uid is expected to be as-
pirated using the ViroSafe® Filter or the BUFFALO FILTER Turbo Auto-Sense™ system, uid
collection devices must be installed with the vacuum hose assembly. Failure to install a uid
collection device may cause lter blockage and/or electrical damage.
Filter Removal Instructions:
1. Aer the ViroSafe® Filter has been exhausted and requires changing, turn the smoke
evacuation system o and disconnect any accessory tubing attached to the lter.
2. Pull the ViroSafe® Filter from the smoke evacuation system and dispose of in accordance with
hospital policy. e ViroSafe® Filter may be disposed of or incinerated.
3. Clean the unit with appropriate germicide prior to re-use. Follow the indicated instructions
for maintenance and installation of a new ViroSafe® Filter.
CAUTION: Using any other lter or accessory not supplied by BUFFALO FILTER may cause
damage to the system and/or cause the system to be inoperable and may void the warranty.
WARNING: e ViroSafe® Filter should be changed when the Filter Life Indicator on the
LCD displays “REPLACE”. Failure to change this lter may result in decreased eciency and
contamination of the electric motor, vacuum pump, and sound absorbing media within the
system, or non-operation of smoke evacuator.
Bualo Filter P/N: PSWTURBO AS & PSWTURBO AS/220
ViroSafe® Filter P/N: (Call for selection)
Conguration: Portable or tabletop
ViroSafe® Filters - Multi Port Filter: 4-Stage Filtration In One Casing,
(Pre-Filter, ULPA, Carbon, Post-Filter)
Filter(s): ULPA
Particle Size, µm: 0.1 to 0.2 Microns At 99.999% Eciency
Filter Life: Automatic Factory Set Filter Sensor
Filter Life Indicator: Timed Replacement