Instruction Manual

Bualo Filter Turbo Auto-Sense™ Operators Manual: Page 5
10. is equipment needs special precautions regarding Electromagnetic Compatibility and
needs to be installed according to EMC information found in this manual.
11. is equipment utilizes mobile RF communications equipment that can aect medical elec-
trical equipment.
12. is equipment has been tested and found to comply with the limits for a Class A digital de-
vice, pursuant to Part 15 of the FCC rules. ese limits are designed to provide reasonable
protection against harmful interference when the equipment is operated in a commercial
environment. is equipment generates, uses, and can radiate radio frequency energy and,
if not installed and used in accordance with the instruction manual, may cause harmful
interference to radio communications. Operation of this equipment in a residential area is
likely to cause harmful interference in which case the user will be required to correct the
interference at their expense.
13. is equipment operates in the following radio frequency specications:
RX modulation: Pulse-width coded, AM 100% modulation
TX Frequencies: Manchester encoded,
A = fc =/- 423.75kHz, B = fc +/- 484.29kHz
Low bit: transition A to B
High bit: transition B to A
14. To isolate equipment from supply mains, unplug the power cord from the appliance inlet on
the unit or receptacle in the wall. Position the equipment to allow for ease of unplugging
power cord.
15. Potential Equalization Conductor: Terminal located on back panel for connection of poten-
tial equalization. Conductor complies with requirements per IEC 60601-1 (2005).
e Turbo Auto-Sense™ Smoke Evacuation System(s) and all lters are not intended for con-
tact with patients.
1.4 Cautions and Warnings
Please note that all Cautions and Warnings should be read and understood before any use of this
equipment.
Please note that all Cautions and Warnings
should be read and understood before any
use of this equipment.
1.4.1 WARNINGS:
• Read this manual thoroughly, and be familiar with its contents prior to using this equipment.