22Bualo Filter® VisiClear™ Operator’s Manual
2.5 Electromagnetic Compatibility Information per IEC60601-1-2
Table 1
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
e Surgical Smoke Plume Evacuator model VisiClear™ is intended for use in the electromagnetic environment
specied below. e customer or user of the VisiClear™ should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1 e VisiClear™ uses RF energy only for its internal function.
erefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class A e model VisiClear™ is suitable for use in all establishments,
other than domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic
Emissions
IEC 61000-3-2
Class A Not applicable.
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Class A Not applicable.