2.5 Electromagnetic Compatibility Information per IEC60601-1-2
Table 1
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
e Surgical Smoke Plume Evacuator model VisiClear™ is intended for use in the electromagnetic environment
specied below. e customer or user of the VisiClear™ should assure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
Group 1 e VisiClear™ uses RF energy only for its internal function.
erefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class A e model VisiClear™ is suitable for use in all establishments,
other than domestic establishments and those directly
connected to the public low-voltage power supply network
that supplies buildings used for domestic purposes.
Harmonic
Emissions
IEC 61000-3-2
Class A Not applicable.
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Class A Not applicable.