User Manual

27Bualo Filter® VisiClear™ Operators Manual
3.4 Troubleshooting the System – see below.
PROBLEM POTENTIAL CAUSE CORRECTIVE ACTION
1. Smoke Evacuation
System is ON but
suction is minimal
or none.
1. Filter is not seated
completely.
2. Filter is clogged.
3. Vacuum hose or tube is
clogged.
4. Motor/blower is
obstructed.
1. Re-install VisiClear™ Filter, press rmly into place
and fully engage clip.
2. Replace lter with a genuine Bualo Filter
VisiClear™ Filter.
3. Replace vacuum hose or tube with genuine
Bualo Filter products.
4. Call BioMed or BUFFALO FILTER® Technical
Services at 1.800.343.2324 or 716.835.7000.
2. Smoke Evacuation
System does not
function even though
suction ON button is
depressed.
1. Not plugged into an
electrical outlet.
2. Fuses are blown.
3. Electronic system
failure.
4. Filter life has expired or
invalid lter inserted.
1. Check power outlet and connection to rear or side
panel of the machine.
2./3. Call BioMed or BUFFALO FILTER® Technical
Services at 1.800.343.2324 or 716.835.7000.
4. Replace lter with a genuine Bualo Filter
VisiClear™ Filter.