User Manual

12
Start-up
8.3 Dosing quantity
The dosing unit is set to a pumping volume of 5µl/pump stroke as
standard. This value cannot be changed.
8.4 Selecting the dosing mode
By default, the dosing unit is configured so that it doses at a fre-
quency of 5 Hz once the power supply voltage is connected. The
dosing mode can be pre-selected on the rotary switch for the elec-
tronics once the cover has been removed (see “Fig. 6: Setting the
dosing mode”).
DANGER!
Risk of electric shock!
Before reaching into the device or equipment, switch off the power
supply and secure to prevent reactivation!
Only adequately trained personnel may reach into the device!
Rotary switch for pre-selecting the
dosing mode
Fig. 6: Setting the dosing mode
Making adjustments using the rotary switch for the electronics:
0 Impulse mode
>120 ms
Unit doses precisely 5 µl
(= 1 stroke) and then stops.
Power supply voltage must be at
least 120 ms for this purpose. If the
power supply is then interrupted
for > 2 ms and is subsequently
reconnected, a stroke is initiated, etc.
1 40 Hz frequency mode Unit is in frequency mode and
doses at the selected frequency
once the power supply voltage is
connected.
2 25 Hz frequency mode
3 10 Hz frequency mode
4
5 Hz frequency mode
(standard)
5 5 Hz service mode;
unit stops after
100 cycles
For servicing purposes: Following
connection of the power supply
voltage, the unit operates at a fre-
quency of 5 Hz for precisely
100 pumping cycles and then
stops.
The supply rate can be checked
using this function.
english
Type 7615