User Manual

14
Technical data
Fluid temperature, compact version
8025, compact version
8035
The fluid temperature may be restricted by the fluid
pressure and the materials the fitting used is made
of. See “Figure 7”, chap. “7.2.1” or “Figure 14”, chap.
“7.5”.
-15 to +80 °C
-15 to +100 °C
Fluid pressure, compact version
8025, compact version
8035
The fluid pressure may be restricted by the fluid
temperature and the materials the fitting used is made
of. See “Figure 7”, chap. “7.2.1” or “Figure 14”, chap.
“7.5”.
PN10
PN10 or PN16
Flow rate measurement, compact version
Measurement range
measurement error, with a K factor Teach-In procedure
Measurement error, with K factor of the fitting used
Linearity
Repeatability
0,3 to 10 m/s
• ±0,5 % of the full scale
1)
±(0,5 % of the full scale + 2.5 % of the measured
value)
1)
±0,5 % of the full scale
1)
±0,4 % of the measured value
1)
Determined in the following reference conditions: fluid = water, water and ambiant temperatures = 20 °C, upstream and
downstream distances respected, appropriate pipe dimensions.
6.3.2 Mechanical data
Part Material
Housing and cover with lid, compact version PC, lid with UV filter
Housing and cover, wall-mounted version ABS
Cable glands, compact or wall-mounted versions PA
Open housing, panel-mounted version PC
Foil Polyester
4 screws Stainless steel
Flow sensor holder, 8025 compact version PVDF
Axis and bearing of the paddle-wheel, 8025 compact
version
ceramics
Paddle wheel, 8025 compact version PVDF
O-ring seal, 8025 compact version FKM (or EPDM, delivered with the device)
Nut, 8025 compact version PC
Specifications of the fitting refer to the related operating instructions
Cable clips PA
English
Type 8025 - 8035 BATCH