Installation and wiring
7.6.4 Wiring a compact version
Only move the selectors when the power supply is off.
Insert the supplied stopper gasket into the unused cable gland to ensure the tightness of the device.
• Unscrew the unused cable gland.
• Remove the transparent disk.
Insert the supplied stopper gasket.
• Screw the nut of the cable gland.
Set the selector "SENSOR TYPE" on "NPN/PNP". See “Figure 22”.
Never set selector "SENSOR TYPE" on the left on a compact version.
Selector
C
makes it possible to configure the type of signal received from the flow sensor.
On a compact version, set the selector on the right (default position).
C
SENSOR TYPE
COIL NPN/PNP
SENSOR TYPE
COIL NPN/PNP
Figure 22: Using selector "SENSOR TYPE" on a compact version
Set selector "SENSOR SUPPLY" depending on the value of the power supply of the dosing controller. See
“Table 4”.
Set selector "LOAD": See “Table 4”.
English
Type 8025 - 8035 BATCH