7. QUICK INSTALLATION
Contact your Bürkert
agent
→ Unpack the device.
Order code
correct?
→ Loosen 4 screws and remove
cover.
yes
No
→Select the frequency of the main
supply with switch 1, see chap.
9.4.
→ Select the filter with switches 2
and 3, see chap. 9.5.
→ Select the current output type,
see chap. 8.3.1.
The following formulae are used to calculate the K
1
and K
2
factors
needed to convert the current or frequency into a flow rate:
Full scale K factor K
1
K factor K
2
10 m/s
K
1
=
100
K
fitting
K
2
=
20
3*K
fitting
5 m/s
K
1
=
200
K
fitting
K
2
=
40
3*K
fitting
2 m/s
K
1
=
500
K
fitting
K
2
=
100
3*K
fitting
where K
fitting
= K factor of the S020 fitting used
Example:
If the full scale of the device is set to 5 m/s, the value of the current
output will be:
I = 40 Q + 4
3*K
fitting
with I in mA, K
fitting
in pulse/litre and Q in l/s.
English
Type 8041