Manual

20
INstallation
Vertical mounting
Correct Incorrect
flow direction
Fig. 10: Filling of the pipe
Correct
flow direction
Incorrect
Correct Incorrect
Fig. 11: Air bubbles within the pipe
If necessary, use a flow conditioner to improve measurement
precision.
8.2.2. Installation into the pipe of a 8041
with a G2'' nut
Observe the installation recommendations described at
chap. 8.2.1 and in the operating instructions of the S020.
Install the S020 fitting on the pipe.
Insert the nut (see mark 3, Fig. 12) on the fitting.
Insert the snap ring (mark 2, Fig. 12) into the groove (mark 5,
Fig. 12).
Position the cable glands parallel to the pipe and insert the
device (mark 1, Fig. 12) into the fitting.
Tighten the nut (mark 3, Fig. 12) by hand on the device.
1
2
3
5
4
Electrodes
Flow direction
Fig. 12: Installation into the pipe of the flowmeter with a G2'' nut
English
Type 8041