23
INstallation
note
The device is not tight if only one or none of the cable glands
is used
The device is only tight when the cable glands are either wired or
sealed. To seal a cable gland, do the following:
Loosen the nut of the unused cable gland.
Remove the transparent disk.
Insert the supplied stopper gasket.
Screw the nut of the cable gland.
Loosen the nuts of the cable glands.
Insert each cable through a nut than through a cable gland.
123456
Power cable shield
Earth cable coming from the
housing
Second cable on a version with
a flow sensor holder in stainless
steel
Fig. 15: Earth connection terminal
+
-
18-36 VDC
1
2
2
Power cable shield
Power supply
Fig. 16: Earthing the device
Wire the 4-20 mA current output (see 8.3.1).
Wire the frequency output (see 8.3.2).
Wire the relay output (see 8.3.3).
Put the cover of the housing as described in Fig. 17.
Screw the 4 screws in an alternating pattern.
Fig. 17: Position of the cover of the device