38
Adjustment and commissioning
Calibration in progress
Press and hold the push
button until...
To select the calibration of the
other parameter
B
A
Briefly press push button.
End of calibration: red LED
goes off?
Go back to Read mode
NO
To calibrate automatically
Red LED flashes:
4 times: fluid velocity < 2 m/s
5 times: fluid velocity < 10 m/s
The full scale cannot be calibrated
Press push button for 2 s.
Go back to Read mode using the
full scale selected or previously
calibrated.
Full scale value stored
YES
Wait for 10 s.
To go back to the Read mode
C
Go back to Read mode without con-
firming the selection displayed.
...bargraph goes back to step
1, Fig. 32
Fig. 33: Calibration of the full scale, part 2
English
Type 8041