Maintenance and troubleshooting
note 
After cleaning of the flow sensor:
• Rince the flow sensor.
• Check the seals and replace them if necessary.
• Before commissioning calibrate the flow zero point.
To prevent any measurement error due to deposits on the electrodes, clean the wetted parts regularly (cleaning 
frequency depends on the process).
9.4.  Replacing the seal on a 8045 with G2" nut
note 
Do not scratch the seal groove.
1
2
3
Seal 
Groove 
Fig. 62:  Dismounting of the flowmeter and location of the seal
 → Loosen the nut of the flowmeter (mark 2).
 → Remove the flowmeter (mark 1) from the fitting.
 → Remove the seal from the groove.
 → Clean the seal groove.
 → Insert the new O-ring seal in the groove (see chap. 10).
 → Insert the flowmeter into the fitting.
 → Tighten the nut (mark 2) by hand on the flowmeter.