Instruction Manual

27
Installation and wiring
Correct Incorrect
Figure 15 : The oval gear axes must be horizontal (seen from the front)
Fit the display module (see chap. 7.4) to parameter the device.
Assemble flowmeter type 8076 (see Figure 13)
Set the K factor parameter (see chap. 9.13.5) or have it determined through a teach-in procedure.
8.3. Wiring
danger
Risk of injury due to electrical voltage.
• Shut down and isolate the electrical power source before carrying out work on the system.
• Observe all applicable accident protection and safety regulations for electrical equipment.
Use a high quality electrical power supply (filtered and regulated).
• Make sure the installation is equipotential. See chap. 8.3.3.
Use a shielded cable.
• Once the device is wired, set the "HWMode" parameter depending on the wiring carried out, sink/NPN
or source/PNP. See chap. 9.12.8.
8.3.1. Electrical connections
Number of fixed connectors Type of connectors
1 male M12 fixed connector
female 5-pin M12 (available as an accessory: see chap. 11)
1 male M12 fixed connector and 1 female
M12 fixed connector
female 5-pin M12 + male 5-pin M12 (both available as accessories:
see chap. 11)
8.3.2. Assembling the male or female connector (accessories:
4 3 2 1
Unscrew the nut [1] on the body [4].
Insert the cable into the nut [1], the cable clamp [2] and the seal [3], and then into
the body [4].
English
Type 8026- 8036- 8076