Adjustment and commissioning
 → Run the teach-in:
 → Open the valve to fill the tank.
 → Save the new K 
factor calculated or 
not.
Vol.Teach
Sensor
Start Teach
Calib
K teached:
K=3,810
Cal. Result
BACK
Save: Yes/No
Cal. Result
The device displays the K factor 
calculated by teach-in.
OK
The device displays the instantaneous 
flow of the fluid.
OK
K teached set to 
max K=9999
BACK
 → Confirm the K factor =9999 with “YES” 
or keep the current K factor with “NO”.
If the calculated K factor is 
≥ 9999, it is forced to 9999.
Meas. Progr.
 → When the tank is filled, close the valve.
OK
2.001 l/s
The device displays the volume calcu-
lated with the current K factor. If the 
display shows “++++”, the volume 
calculated > 9999 volume units: conti-
nue the procedure as normal. 
 → Enter the actual volume of fluid 
in the tank.
Input Vol.
101.2 l
+099.0 l
Save modified 
data? Yes/No