Fig. 3: Adjustment of the angle of inclination
Fig. 4: Turning by fastening in the centre
4.4 Instructions for installation
For tight installation of the version with plastic horn antenna with
compression or adapter ange, the following conditions must be
fullled:
1 Use suitable at seal, e.g. of EPDM with Shore hardness 25 or 50
2 Make sure the number of ange screws corresponds to the
number of ange holes
3 Tighten all screws with the torque stated in the technical data
The emitted radar impulses of the radar sensor are electromagnetic
waves. The polarisation plane is the direction of the electrical wave
component. By turning the instrument in the connection ange or
mounting boss, the polarisation can be used to reduce the eects of
false echoes.
The position of the polarisation plane is marked on the process tting
of the instrument.
Ti
ght installation of the
plastic horn antenna
Polaris
ation plane
12 LEVEL TRANSMITTER 8136 4 20 mA/HART - two-wire
4 Mounting
41783-EN-120316