Manual
Menu Menu
item Default value
Lock operation Released
Display Language Like order
Displayed value Distance
Display unit m
Scaling size Volume
l
Scaling 0.00 lin %, 0 l
100.00 lin %, 100 l
Lighting Switched off
Further settings Distance unit m
Temperature unit °C
Probe length Length of the standpipe Ex factory
Linearisation
curve
linear
HART mode Standard
Address 0
6.4 Saving
the parameter adjustment data
We recommended noting the adjusted data, e.g. in this operating
instructions manual, and archiving them afterwards. They are thus
available for multiple us e or service purposes.
If the instrument is equipped with an indicating and adjustment
module, the data in the sensor can be saved in the indicating and
adjustment module. The procedure is described in the operating
instructions manual "Indicating and adjustment module" in the menu
item "Copy sensor data". The data remain there permanently even if
the sensor power supply fails.
The following data or settings for adjustment of the indicating and
adjustment module are saved:
l All data of the menu "Setup" and "Display"
l In the menu "Additional settings" the items "Sensor-specific units,
temperature unit and linearization"
l The values of the user programmable linearisation curve
The function can also be used to transfer settings from one instrument
to another instrument of the same type. If it is necessary to exchange a
sensor, the indicating and adjustment module is inserted into the
replacement instrument and the data are likewise written into the
sensor via the menu item "Copy sensor data".
44 LEVEL TRANSMITTER 8136 • 4 … 20 mA/HART - two-wire
6 Set up with the indicating and adjustment module
41783-EN-120316










