Manual
- Alum
inium/Stainless steel housing 50 Nm (36.88 lbf ft)
Input variable
Measured variable The measured quantity is the distance between
process fitting of the sensor and product surface.
The reference plane is the seal surface on the
hexagon or the lower side of the flange.
3
4
2
1
Fig. 41: Data of the input variable
1 Reference plane
2 Measured variable, max. measuring range
3 Antenna length
4 Useful measuring range
Standard electronics
Max. measuring range 35 m (114.8 ft)
Recommended measuring range
- Encapsulated antenna system up to 10 m (32.81 ft)
- Plastic horn antenna up to 20 m (65.62 ft)
Electronics with increased sensitivity
Max. measuring range 35 m (114.8 ft)
Recommended measuring range
- Encapsulated antenna system up to 10 m (32.81 ft)
- Plastic horn antenna up to 20 m (65.62 ft)
Output variable
Output signal 4 … 20 mA/HART
Fulfilled HART specification 7.0
60 LEVEL TRANSMITTER 8136 • 4 … 20 mA/HART - two-wire
11 Supplement
41783-EN-120316