Installation and wiring
7.3.6.  Wiring the "COND" conductivity module
Conductivity 
sensor
C+
P+
P-
C-
GD
FE
SE
TS
TS
1 2 3 4 5 6 7 8 9
Temperature 
sensor
FE = functional earth
Figure 18 :  Wiring a resistive conductivity cell with 2 electrodes and a Pt100 or Pt1000 temperature sensor in a conduc-
tivity module
C+
P+
P-
C-
GD
FE
SE
TS
TS
1
3
2
8220
1 2 3 4 5 6 7 8 9
FE = functional earth
Figure 19 :  Connection example for a 8220 conductivity sensor
C+
P+
P-
C-
GD
FE
SE
TS
TS
1 2 3 4 5 6 7 8 9
Temperature 
sensor
FE = functional earth
Figure 20 :  Wiring a resistive conductivity cell with 4 electrodes and a Pt100 or Pt1000 temperature sensor in a conduc-
tivity module
Type 8619