User guide
35
Adjustment and commissioning
Calibration
Cond manual calibMx:Conductivity
2
4.294 S/cm
25.01 °C
0.000
µS/cm
Cond. calib.
Move the decimal point by 
pressing 
 to the far 
right of the numerical value 
and then on 
 until the 
decimal point is in the desired 
place (the decimal point 
moves in a loop)
Increment or decrement the figure selected by 
pressing 
 or 
After confirming the entered 
numerical value by pressing 
"OK", modify the unit selected by 
pressing 
 or 
Figure 25 :  Example of a numerical value entering
• Access, for example, the pH value simulation function. Refer to chap. 8.9 to access "Tests" menu.
Tests
PV:
Value:
Simulate PV M1:pH/ORP
ENTERING
pH
T
- 1.000
pH
Simulation value
Select the character to the far left by pressing 
 and then allocate the "+" or "-" sign 
by pressing
.
To exit the "Tests" menu, press the dynamic key, "ABORT"
Figure 26 :  Changing the sign of a numerical value
Type 8619
english