Adjustment and commissioning
Graph 
zone
Description Name displayed in 
the user defined 
"Ux" view (see 
chap. 8.10.8)
Associated 
code in the 
datalogger (see 
chap. 8.10.18)
State of 
the "on/off" 
output (see 
chap. 8.10.22)
B The conductivity of the fluid is between the 
percentage set in the function "USP Alarm" 
and the value in the table USP<645> at 
the corresponding temperature.
"USP Alarm" 2 ON (output not 
inverted)
A The conductivity of the fluid is below the 
percentage set in the function "USP Alarm" 
at the corresponding temperature.
"OK" 0 OFF (output 
not inverted)
TEMP. CALIB: Choose the value of the temperature used when calibrating the sensor:
• choose "Auto": the temperature of the fluid is measured by the sensor
•  choose "Manual": enter the value of the temperature (in °C) in the next field, e.g. when no temperature sensor is 
connected to the module.
CALIBRATION BUFFER: Choose the calibration solution used for the automatic calibration of the conductivity 
sensor.
8.11.  Calibration menu
8.11.1.  Activate/deactivate the Hold function
Refer to chap. 8.9 to access Calibration menu.
Calibration System Hold
Disable
Enable
The Hold mode is automatically deactivated when the multiCELL restarts after a power interruption, if the 
Hold mode was activated at the moment of the power cut-off.
The Hold mode is used to carry out maintenance work without interrupting the process.
To activate the HOLD mode:
 → access the "HOLD" function;
 → choose "Enable";
 → confirm by "OK".
When the device is in Hold mode:
• the display shows the icon 
H
 instead of the icon  ;
•  the current emitted on each 4-20 mA output is fixed at the last value of the process input associated with each 
output;
•  each digital output is fixed at the state acquired at the moment the Hold function is activated;
•  the multiCELL remains in Hold mode until the Hold function is deactivated.