User guide
89
Adjustment and commissioning
• VOLUME TEACH: Determine the K factor unique to your installation using a teach-in procedure by volume. 
Follow the procedure below.
• FLOW TEACH: Determine the K factor unique to your installation using a teach-in procedure by flow rate. 
Follow the procedure on next page.
Detailed procedure for teach-in by volume
 → Prepare a tank capable of containing 100 litres, for example;
 → Choose the volume unit and the flow rate unit in which the teach-in is run:
Mx:Inputs
Calibration
DI1/2: FlowM0:Inputs
Volume teaching L
m3
gal
Igal
Volume unit
L/h
Igal/s
Flow unit
...
 → Running teach-in by volume:
 → Open the valve to fill the tank.
The multiCELL displays the K factor calculated 
by teach-in.
OK
The multiCELL displays the 
instantaneous flow rate of the fluid.
OK
 → When the tank is filled, close the valve.
OK
2.001 l/s
The multiCELL displays the 
volume calculated with the 
current K factor. 
 → Enter the actual volume of fluid in 
the tank.
101.2 l
+099.0 l
Volume teaching
Calibration
DI1:/DI2:FlowM0:Inputs
Start teaching
Meas. in progr.
Flowed volume
Flow K factor:
K=3,810
Teaching result
Type 8619
english