Adjustment and commissioning
The multiCELL displays the calibration result.
1)
Rinse
7.001 pH
-59.6 mV
25.00 °C
Immerse the clean probe in the first
buffer solution: the multiCELL displays
the measured pH of the solution.
Enter the pH of the buffer solution (indicated on the bottle)
Wait until the measured pH stabilises.
Calibration Mx:pH/ORP pH manual calib.
OK
F4
1st point
7.035 pH
7.000 pH
pH Calibr. result
Zero: 7.000 pH
Slope:-59.15mV/pH
Calibrate a 2nd
point?
NO
Rinse the probe.
confirm rinsing by "OK" when the pH
reaches the desired value.
OK
F4
2nd point
2.967 pH
2.000 pH
Immerse the clean probe in the second buffer solution:
the multiCELL displays the measured pH of the solution.
Enter the pH of the second buffer solution (indicated on the
bottle)
Wait until the measured pH stabilises.
OK
F4
pH Calibr. result
Zero: 7.000 pH
Slope:-59.15mV/pH
YES
The multiCELL displays the
calibration result.
1)
1)
a possible "warning" message indicates either an error in the buffer solution or the ageing of the probe.
a possible "error" message indicates that the probe must be replaced.