The controller 8039 can be cleaned with water or any solution compatible
with the materials the device is made of.
For more information, please contact your Bürkert sales office.


  
ERR 0 Calibration data is lost.
Reading error: the process is stopped.
Press the ENTER key to go back to the Normal mode.
The device has returned to its default configuration: the device
must be calibrated again.
If the message appears frequently, send the device back to your
Bürkert sales office.
ERR 1 Calibration data cannot be saved.
Write error: the process is stopped.
Press the ENTER key to go back to the Normal mode.
The device displays the configured data; BUT this data has not
been saved: the device must be calibrated again.
If the message appears frequently, send the device back to your
Bürkert sales office.
ERR 2 The calibration parameters cannot be accessed.
Menu reading error: the process goes on ope-
rating.
Press the UP and DOWN keys under the display to scroll through
the menus.
If the message appears frequently, send the device back to your
Bürkert sales office.