130258 GAXT QRG (D5562-6-EN).book Page 0 Monday, June 21, 2010 9:55 AM Limited Warranty and Limitation Liability BW Technologies LP (BW) warrants the product to be free from defects in material and workmanship under normal use and service for a period of two years, beginning on the date of shipment to the buyer. This warranty extends only to the sale of new and unused products to the original buyer.
GasAlert Extreme Introduction Safety Information - Read First This quick reference guide provides basic information to operate the GasAlert Extreme gas detector. For complete operating instructions, refer to the GasAlert Extreme User Manual provided on the CD-ROM. The GasAlert Extreme gas detector (“the detector”) is designed to warn of hazardous gas levels above user defined alarm setpoints.
GasAlert Extreme Quick Reference Guide • • • • • 2 Deactivating the detector resets the time-weighted average (TWA), short-term exposure limit (STEL), and maximum gas exposure values to 0. If the detector is damaged or parts are missing, contact BW Technologies by Honeywell immediately. Do not use the detector if it is damaged. Before using the detector, inspect the case. Look for cracks and missing parts.
GasAlert Extreme Parts of the GasAlert Extreme Parts of the GasAlert Extreme Item Description 1 Visual alarms bars (LEDs) 2 Liquid crystal display (LCD) 3 Pushbuttons 4 Audible alarm 5 Sensor and sensor grill 6 Infrared (IR) communication port 7 Alligator clip 3
GasAlert Extreme Quick Reference Guide Display Elements Item Note Press and hold (until the backlight activates) any button to activate the backlight for 6 seconds. When the backlight option is enabled, the backlight automatically activates for 3 seconds whenever there is insufficient light to view the LCD. The detector is shipped with the backlight option enabled. The backlight does not activate when stealth mode is enabled.
GasAlert Extreme Pushbuttons Pushbuttons Pushbutton Description • To activate the detector, press A. A • To deactivate the detector, press A and hold for 5 seconds. • To enable/disable the confidence beep, while the detector is deactivated press and hold C. While holding C, press A to enable or disable the confidence beep during startup. • To decrement the displayed value or to scroll down, press E.
GasAlert Extreme Quick Reference Guide Calibration Procedure Screen Procedure 1. In a safe area free of hazardous gas and in an atmosphere of 20.9% oxygen, press E and C simultaneously and hold for 5 seconds. The detector beeps and vibrates four times. After CAL. displays on the LCD, the detector then beeps once more. 5. When the LCD flashes , attach the calibration cap and apply gas at a flow rate of 500 to 1000 ml/min. The detector beeps three times at the end of the span stage.
GasAlert Extreme Attach the Gas Cylinder to the Detector Attach the Gas Cylinder to the Detector Item Description 1 Calibration cap 2 Hose 3 Regulator 4 Gas cylinder 7
GasAlert Extreme Quick Reference Guide Alarms Refer to the following table for information about alarms and corresponding screens. During an alarm condition, the detector activates the backlight and the LCD displays the current ambient gas reading.
GasAlert Extreme Alarms Alarm Screen Alarm Automatic Shutdown Alarm (Low battery) • Eight beeps, flashes, and vibrations • LOW displays After Automatic Shutdown (Low battery) • No beep Confidence Beep • One beep every 5 seconds • One quick vibration per minute • No flash or vibrations • Screen Automatic Shutdown Alarm (Calibration past) • Eight beeps, flashes, and vibrations Note displays for a short time If the detector enters low battery alarm, the confidence beep deactivates.
GasAlert Extreme Quick Reference Guide User Options Menu Note When selecting a user option, Set flashes and the LCD displays the opposite of what is currently enabled. To access the user options menu, press and hold E and D simultaneously until the LCD displays OPTN. To scroll through the options, press E or D. Press C to select the option. The following are the available user options: • EXIT: Exits the user options menu. • CLCK (Clock): Define the date and time of the detector.
GasAlert Extreme Maintenance Maintenance To keep the detector in good operating condition, perform the following basic maintenance as required: • Calibrate, test, and inspect the detector at regular intervals. • Keep an operations log of all maintenance, calibrations, and alarm events. • Clean the exterior with a soft damp cloth. Do not use solvents, soaps, or polishes. • Do not immerse the detector in liquids.
GasAlert Extreme Quick Reference Guide 4. Insert the new battery. 5. Re-assemble the detector. When assembling the detector be aware of the following: • • • Clean the seal on the front and rear shells with a soft damp clean cloth. Do not use solvents, soaps, or polishes. Refer to Internal Seals of the GasAlert Extreme. Ensure the front and rear shells are properly aligned to ensure a proper environmental seal.
GasAlert Extreme Maintenance 4. Remove the PCB. Place the PCB on a clean, nonconductive surface. 5. Remove the screen. 6. Using a soft, clean brush, wash the screen with clean, warm water. 7. Insert the sensor screen with the shiny side facing the sensor grill. Internal Seals of the GasAlert Extreme Note Ensure the screen is dry before reinserting into the detector. 8. Re-assemble the detector.
GasAlert Extreme Quick Reference Guide Internal Parts of the GasAlert Extreme Item 14 Description 1 Detector rear screws 2 Rear shell 3 Battery 4 Main PCB screws 5 Main PCB 6 Sensor 7 Sensor screen 8 Front shell
GasAlert Extreme WEEE Directive and Battery Directive WEEE Directive and Battery Directive Specifications Failure to comply with the following battery removal and disposal instructions may result in battery shorting, battery leakage, and/or other damage. Ensure a qualified technician completes the following procedures. Instrument dimensions: 2.8 x 5.0 x 9.5 cm (1.1 x 2.0 x 3.75 in.) Weight: 82 g (2.9 oz.
GasAlert Extreme Quick Reference Guide GasAlert Extreme SO2: 0-150.0 ppm (0.1 ppm increments) GasAlert Extreme Cl2: 0-50.0 ppm (0.1 ppm increments) GasAlert Extreme NH3: 0-100 ppm (1 ppm increments) GasAlert Extreme NH3 (high range): 0-400 ppm (1 ppm increments) GasAlert Extreme NO2: 0-100.0 ppm (0.1 ppm increments) GasAlert Extreme HCN: 0-30.0 ppm (0.1 ppm increments) GasAlert Extreme ETO: 0-100.0 ppm (0.1 ppm increments) GasAlert Extreme ClO2: 0-1.00 ppm (0.01 ppm increments) GasAlert Extreme O3: 0-1.
GasAlert Extreme Specifications This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules and ICES-003 Canadian EMI requirements. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
GasAlert Extreme Quick Reference Guide 18
GasAlert Extreme Specifications 19
GasAlert Extreme Quick Reference Guide 20