Activate and Deactivate the Detector
Only activate the detector in a normal atmosphere (20.9% O
2
) that is free of hazardous
gas.
Hand-aspirated remote sampling only provides continuous gas readings as long as the
bulb is being operated.
BW Technologies by Honeywell recommends that the detector be calibrated at least
once every 6 months or in accordance with customer site procedures, whichever is
sooner. Accurate detection of gases by the detector should be confirmed with test gas
of known concentration before each use.
Use of the ImpactXtreme Enforcer accessory is strongly recommended as it enables
calibration to be performed quickly and easily.
A sensor which cannot be calibrated or which is found to be out of tolerance should be
replaced immediately.
Before You Begin
Each time the detector is activated, it automatically performs a series of self-tests to confirm that
it is fully functional. During testing, you may be prompted to acknowledge messages or follow
test procedures.
1.
Move to a normal atmosphere (20.9% O2) that is free of hazardous gas.
2.
Inspect the detector and accessories for damaged or missing parts. If parts are damaged or
missing, contact BW Technologies by Honeywell or an authorized distributor immediately.
3.
Ensure that the battery pack is charged. See Battery Packs for more information.