ImpactXtreme Operator’s Manual 8 March 2013
12. If the last calibration failed and the calibration interval is not zero, the detector displays the
last failed calibration result.
13.
If the calibration interval is not zero, the detector displays the number of days remaining
until the next calibration is due.
14.
If calibration is due and Forced Calibration is enabled, the Forced Calibration screen is
displayed. For more information, see Calibration and User Options and Settings.
a)
Follow on-screen instructions
b)
If a calibration is not performed, the detector automatically deactivates.
15. If the last bump test failed and bump test interval is not zero, the detector displays the last
failed bump test result.
16.
If the bump test interval is not zero, the detector displays the number of days until the next
bump test is due.
17.
If Bump Test is overdue and Forced Bump Test is enabled, the Forced Bump Test screen is
displayed. For more information, see Bump Test and User Options and Settings.