ImpactXtreme Operator’s Manual 8 March 2013
a) When bump test succeeds, is displayed above the sensor channel.
b) When bump test fails, is displayed above the sensor channel. For more information
on failed bump test results, see Troubleshooting.
8)
When prompted, turn off the gas. Disconnect the tubing from the calibration adapter and
regulator. Remove the calibration adapter from the detector.
a)
The detector purges the test gas until readings are zero.
9)
When prompted, select Yes to bump test another sensor, or No to return to the Bump Test
Menu.
10) Press to return to normal operation.