iERP: XXXXXX DXXXX/X [language/language] © BW Technologies 2008. All rights reserved.
Table of Contents Heading Page Introduction ......................................................................................................................................................................... 1 Zeroing the Sensors ........................................................................................................................................................... 1 Contacting BW Technologies by Honeywell ............................................................................
GasAlertMicroClip XT Technical Reference Guide Title Page Last Calibration Failed (optional) .............................................................................................................................. 14 Overdue Calibration .................................................................................................................................................. 14 Cal IR Lock .........................................................................................................
GasAlertMicroClip XT Technical Reference Guide Title Page Auto Zero on Startup...................................................................................................................................................... 25 O2 Auto-Calibration on Startup (Automatic O2 Calibration) ........................................................................................... 26 LEL By Vol CH4 ..............................................................................................................
GasAlertMicroClip XT Technical Reference Guide Title Page Automatic Deactivation Alarm ........................................................................................................................................ 36 Bump Test.......................................................................................................................................................................... 37 Calibration ........................................................................................
GasAlertMicroClip XT Technical Reference Guide Title Page Startup Troubleshooting .................................................................................................................................................. 56 Calibration Troubleshooting ............................................................................................................................................ 57 Replacement Parts and Accessories ........................................................................
GasAlertMicroClip XT Technical Reference Guide Title Page vi
List of Figures Figure Title Page 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. Parts of the GasAlertMicroClip XT......................................................................................................................... 8 Display Elements................................................................................................................................................... 9 Connecting the IR Link .....................................................................................
GasAlertMicroClip XT Technical Reference Guide viii
List of Tables Table Title Page 1. 2. 3. 4. 5. 6. 7. 8. 9. 10. 11. 12. 13. 14. 15. Gases Monitored ................................................................................................................................................... 1 International Symbols ............................................................................................................................................ 5 Parts of the GasAlertMicroClip XT.............................................................
GasAlertMicroClip XT Technical Reference Guide x
GasAlertMicroClip XT Introduction Table 1. Gases Monitored a Warning To ensure personal safety, read Safety Information - Read First and the Cautions before using the detector. Gas Detected Unit of Measure Hydrogen sulfide (H2S) parts per million (ppm) The GasAlertMicroClip XT (“the detector”) warns of hazardous gas at levels above user-defined alarm setpoints. Carbon monoxide (CO) parts per million (ppm) Oxygen (O2) percent by volume (%) The detector is a personal safety device.
GasAlertMicroClip XT Technical Reference Guide Contacting BW Technologies by Honeywell Safety Information - Read First To contact BW Technologies by Honeywell, call Use the detector only as specified in this guide and the operator’s manual, otherwise the protection provided by the detector may be impaired. USA: 1-888-749-8878 Canada: 1-800-663-4164 Europe: +44 (0) 1295 700300 Other countries: +1-403-248-9226 Address correspondence to BW Technologies by Honeywell 2840 – 2 Avenue S.E.
GasAlertMicroClip XT a Cautions a Cautions • Warning: Substitution of components may impair Intrinsic Safety. • Before using the detector, refer to Sensor Poisons and Contaminants. • Warning: For safety reasons, this equipment must be operated and serviced by qualified personnel only. Read and understand the user manual completely before operating or servicing. • Do not use the detector if it is damaged. Inspect the detector before using. Look for cracks and/or missing parts.
GasAlertMicroClip XT Technical Reference Guide a Cautions • For use only in potentially explosive atmospheres where oxygen concentrations do not exceed 20.9% (v/v). • Any rapid up-scaling reading followed by a declining or erratic reading may indicate a gas concentration beyond upper scale limit, which may be hazardous. • Only calibrate the detector in a fresh air environment and in a safe area. • Use only BW approved batteries for the GasAlertMicroClip XT detector. Refer to Specifications.
GasAlertMicroClip XT a Cautions Table 2. International Symbols Symbols Description n Approved to both U.S.
GasAlertMicroClip XT Technical Reference Guide Sensor Poisons and Contaminants Several cleaners, solvents, and lubricants can contaminate and cause permanent damage to sensors. Before using cleaners, solvents, and lubricants in close proximity to the detector sensors, read the following cautions and refer to the lists below. a Caution Use only the following BW Technologies by Honeywell recommended products and procedures: • Use water based cleaners. • Use non-alcohol based cleaners.
GasAlertMicroClip XT Getting Started Getting Started The list below provides the standard items included with the detector. If the detector is damaged or parts are missing, contact the place of purchase immediately. • Sensors: H2S, CO, O2, and combustible (LEL) • Calibration cap and hose The detector is shipped with the sensors and rechargeable battery installed. Battery Replacement: To replace the battery, contact BW Technologies by Honeywell. The battery can only be replaced by the manufacturer.
GasAlertMicroClip XT Technical Reference Guide Parts of the GasAlertMicroClip XT Table 3. Parts of the GasAlertMicroClip XT Item Description 1 IntelliFlash 2 Visual alarm indicators (LEDs) 3 Alligator clip 4 Charging connector / IR interface 5 Pushbutton (C) 6 Carbon monoxide (CO) sensor 7 Hydrogen sulfide (H2S) sensor 8 Oxygen (O2) sensor 9 Combustible (LEL) sensor 10 Audible alarm Liquid crystal display (LCD) Figure 1.
GasAlertMicroClip XT Display Elements Display Elements Table 4. Display Elements Item 1 8 2 7 3 6 5 Description 1 Alarm condition 2 Automatically zero sensor 3 Numeric value 4 Stealth mode 5 Battery life indicator 6 Gas identifier bars 7 Gas cylinder 8 Automatically span sensor 4 Figure 2.
GasAlertMicroClip XT Technical Reference Guide Pushbutton Table 5. Pushbutton Pushbutton Description • To activate the detector press C. • To deactivate the detector, press and hold C until the OFF countdown is complete and the LCD deactivates. • To view the TWA, STEL, and peak (maximum) readings, press C twice. To clear the TWA, STEL, and peak readings, press C when the LCD displays RESET. C • To initiate calibration, deactivate the detector.
GasAlertMicroClip XT Activating the Detector Activating the Detector a Caution Only activate the detector in a fresh air environment and in a safe area. Audible/Visual Test 1. All of the LCD elements display simultaneously as the detector beeps, flashes, vibrates, and activates the backlight. To activate the detector, press C. Self-Test The following startup tests are written as startup performance is intended. If an error occurs, refer to Startup Troubleshooting.
GasAlertMicroClip XT Technical Reference Guide Startup Message Sensor and Power Test 3. If data is entered in the Startup option (25 characters maximum) of Fleet Manager II, that data will display during the startup self-test. To enter a startup message, refer to Detector Identification or the Fleet Manager II Operator’s Manual. 5. The detector then tests the sensors. Alarm Setpoints 4. Next, the TWA, STEL, low, and high alarm setpoints display.
GasAlertMicroClip XT Activating the Detector Automatic Zero and O2 Calibration (optional) 6. Auto-Zero on Startup: If enabled, the H2S, CO, and LEL sensors are automatically zeroed during startup. Each sensor is enabled individually. The auto zero option is enabled for each sensor upon shipment. O2 Auto-Calibration on Startup: If enabled, the O2 sensor is automatically calibrated during startup. The auto zero option is enabled for the O2 sensor upon shipment.
GasAlertMicroClip XT Technical Reference Guide Last Calibration Failed (optional) If any sensor failed the last calibration, CAL FAILURE displays on the screen. Note If calibration is not performed, or C is not pressed within 2 minutes, the detector automatically deactivates. If the Force Calibration When Overdue is disabled, press C to acknowledge the warning. The detector continues with the startup self-tests and then enters normal operation.
GasAlertMicroClip XT Activating the Detector Last Bump Test Failed If any sensor failed the last calibration, CAL FAILURE displays on the screen. A bump test must be performed to enter normal operation. Apply gas to the sensors. Ensure the visual, audible, and vibrator alarms activate. When the gas is removed, the detector briefly remains in alarm until the gas has cleared from the sensors.
GasAlertMicroClip XT Technical Reference Guide Self-Test Pass Battery Test When the detector has passed all startup self-tests, it enters normal operation. The LCD displays the ambient gas readings. The battery is tested when the detector is activated and continuously thereafter. A newly charged battery will typically operate for 10-12 hours. • Battery power is continually displayed during normal operation. If battery power is low, L and flashes.
GasAlertMicroClip XT Deactivating the Detector Deactivating the Detector To deactivate the detector, press and hold C. The detector • performs a sequence of two sirens with alternating flashes, • vibrates, • initiates the deactivation countdown, and • displays OFF. Note If C is released before the countdown is complete, the detector will not deactivate.
GasAlertMicroClip XT Technical Reference Guide Installing Fleet Manager II 5. Click the Login/Logout button. Fleet Manager II is required to configure the detector and sensors. And IR Link is also required. To purchase contact BW Technologies by Honeywell. To install Fleet Manager II complete the following: 1. Install Fleet Manager II using the Fleet Manager CDROM (available with MicroDock II and the IR Link), or download (at no cost) from BW Technologies by Honeywell website: www.gasmonitors.com. 2.
GasAlertMicroClip XT Installing Fleet Manager II 9. When the Device Selection dialog box displays, select GasAlertMicroClip / GasAlertMicroClip XT and click OK. Using Fleet Manager II to Configure the Detector When Fleet Manager II is installed, refer to Table 6., Figure 3., and the following procedures: Table 6.
GasAlertMicroClip XT Technical Reference Guide 1. Activate the detector and wait for the startup tests to complete. 2. Connect the USB cable to the USB port on the computer. 3. Connect the USB cable to the IR Link. 4. Insert the IR link onto the IR interface on the back of the detector. 5. Open Fleet Manager II and access the Sensors tab. Refer to Installing Fleet Manager II. 6. From the Sensors tab, click Retrieve from Device at the bottom of the window.
GasAlertMicroClip XT Detector Identification Detector Identification Hardware/Firmware Revision The Detector Identification section provides information about the detector, current firmware revision, and hardware revision. Data can also be entered (25 characters per line) to display as a startup message on the detector LCD each time it is activated. Hardware/Firmware Revision cannot be altered. The field automatically populates when data is retrieved from the detector.
GasAlertMicroClip XT Technical Reference Guide Sensor Configuration Sensor Disabled Settings for the sensors are configured individually. Enter values or enable/disable options. Refer to Factory Gas Alarm Setpoints for setpoint values. a Warning Use extreme caution when disabling a sensor. The disabled sensor cannot detect and alarm against the applicable gas. Note Depending upon the sensor, the options may vary. 1. Click Retrieve from Device to populate the fields with the current detector settings.
GasAlertMicroClip XT Sensor Configuration 3. Click the Save to Device button located at the bottom of the window. 4. The detector LCD automatically updates. The gas type and sensor readings no longer display on the LCD for the applicable sensor. Calibration Interval a Caution BW recommends that the sensors be calibrated once every 180 days (6 months). Define how often a sensor should be calibrated in the Cal Interval field. A different calibration interval can be defined for each sensor. 1.
GasAlertMicroClip XT Technical Reference Guide Bump Interval High Alarm Define how often a bump check should be performed for each sensor in the Bump Interval field. A different bump interval can be defined for each sensor. Enter the high alarm setpoints for each sensor. Refer to Factory Gas Alarm Setpoints for factory defined alarm setpoints. Applicable to all sensors. 1. Enter the value (1-365 days) for each sensor. 2. Enter 0 to disable the Bump Interval option.
GasAlertMicroClip XT Sensor Configuration 2. Enter the setpoint in the TWA Alarm field. STEL Interval STEL Interval provides protection for workers from over exposure to high concentrations of gas, and is based on used-defined 5-15 minute intervals. When the maximum STEL is reached, the detector alarms to notify the worker. a Caution Follow all safety procedures as defined by your employer.
GasAlertMicroClip XT Technical Reference Guide O2 Auto-Calibration on Startup (Automatic O2 Calibration) When enabled, the O2 sensor is automatically calibrated during the startup self-tests. User Options The user options section provides detector features that can be enabled or disabled. The green checkmark indicates the option is enabled. Click the checkmark to disable the option. The detector is shipped with the O2 Auto-Calibration on Startup option enabled.
GasAlertMicroClip XT User Options Cal Lock (Calibration IR Lock) Confidence Beep When enabled, the sensors can only be calibrated using an infrared (IR) device to ensure calibrations are recorded. The following are IR devices: When enabled, the confidence beep provides continuous audible confirmation that the detector is operating correctly by beeping once every second. • IR Link with Fleet Manager (refer to Calibration Procedure), • MicroDock II base station (refer to the MicroDock II User Manual).
GasAlertMicroClip XT Technical Reference Guide with the IR Link. This option is available only if configuring the GasAlertMicroClip with the base station. Default interval is 15 seconds. Force Bump When Overdue When enabled, if a sensor(s) is past due for a bump test, the sensor(s) must be bump tested immediately, otherwise the detector deactivates. A bump test should be performed regularly to ensure the sensors are responding correctly to gas.
GasAlertMicroClip XT User Options IntelliFlash When enabled, the green LED flashes to provide continuous visual confirmation that the detector is operating correctly. Note IntelliFlash is only applicable to GasAlertMicroClip XT. IntelliFlash automatically deactivates during a low battery alarm, a self-test fail, a calibration fail, a bump test fail, or during an alarm event. To define how often IntelliFlash occurs (1-60 seconds), refer to Confidence Beep and IntelliFlash Interval.
GasAlertMicroClip XT Technical Reference Guide Stealth Mode Language Menu When enabled, the backlight, visual alarms, and audible alarms are disabled. displays continuously on the LCD. The detector can display warnings and notifications in five different languages. Refer to the following illustration. During an alarm, the vibrator activates and readings display on the LCD. The detector is shipped with Stealth Mode disabled. Click a language.
GasAlertMicroClip XT Alarms Alarms Table 7. describes the detector alarms and corresponding screens. During an alarm condition, the detector activates the backlight, audible/visual/ vibrator alarms, and displays the current ambient readings. If more than one type or level of alarm occurs simultaneously, a multi-gas alarm results. If Stealth is enabled, the audible and visual alarms are disabled, and only the vibrator alarm activates.
GasAlertMicroClip XT Technical Reference Guide Alarm Screen Alarm Multi-Gas Alarm • Sequence alternating low and high alarm siren and flash • L and gas bars flash • Vibrator alarm activates Over Limit (OL) Alarm • Fast siren and alternating flash • L and gas bar flash • Vibrator alarm activates • OL displays Sensor Alarm • During startup sequence Error [sensor name] displays • During normal operation Err displays (must be acknowledged by press C) Confidence Beep and IntelliFlash • One beep and flash e
GasAlertMicroClip XT Alarms Computed Gas Exposures Viewing Gas Exposures a Warning To avoid possible personal injury, do not deactivate the detector during a work shift. TWA, STEL, and MAX readings reset once the detector is deactivated. To view the TWA, STEL, and peak (maximum) readings, press C twice. The LCD first displays the TWA gas exposures. Table 8.
GasAlertMicroClip XT Technical Reference Guide Clearing Gas Exposures Gas Alarm Setpoints a Caution Follow all safety procedures as defined by your employer. Gas alarms are activated when detected gas concentrations are above or below the user-defined setpoints. Gas alarms are described below. Confirm with your supervisor before clearing TWA and STEL alarms. Table 9. Gas Alarm Setpoints Alarm To clear the TWA, STEL, and peak exposure readings, press C when the LCD displays RESET.
GasAlertMicroClip XT Alarms Factory Gas Alarm Setpoints Changing Alarm Setpoints Note Standard factory alarm setpoints may vary by region. To change alarm setpoints, use the base station or IR Link and refer to the following under Sensor Configuration: • Low Alarm Table 10. lists the factory alarm setpoints as defined by Occupational Safety and Health Association (OSHA). Table 10. Sample Factory Alarm Setpoints Gas TWA STEL • High Alarm • TWA Alarm • STEL Alarm Low High O2 N/A N/A 19.5% vol.
GasAlertMicroClip XT Technical Reference Guide Sensor Alarm The detector tests for missing or defective sensors during the startup self-test and continuously thereafter. • During start-up if a sensor fails, Error and the sensor name displays. • If a sensor fails the self-test, Err displays above the gas type of the failed sensor. The detector will enter alarm. Press C to acknowledge the sensor alarm and stoop the alarm. Refer to Troubleshooting.
GasAlertMicroClip XT Bump Test Bump Test 37
GasAlertMicroClip XT Technical Reference Guide Calibration • Allow the detector to stabilize for 1 minute after activation before performing a calibration or bump test. Guidelines • If a certified calibration is required, contact BW Technologies by Honeywell. When calibrating the detector, adhere to the following guidelines: • Recommended gas mixture: CO: 100 ppm balance N2 H2S: 25 ppm balance N2 LEL: 50% LEL or 2.5% by vol.
GasAlertMicroClip XT Calibration Connecting the Gas Cylinder to the Detector Refer to the following Figure 8., Table 11., and procedures to connect the gas cylinder to the detector for calibration. Note Wind currents may cause false readings and poor calibrations. Table 11. Connecting the Gas Cylinder to the Detector Item 1 Description Calibration cap 2 Calibration hose 3 Gas cylinder with 0.5 ml/min regulator Read the following steps (1-7) before beginning calibration. 1.
GasAlertMicroClip XT Technical Reference Guide Calibration Setup The following calibration procedures are written as calibration performance is intended. If an error or failure occurs, refer to Calibration Troubleshooting. 3. Insert the IR Link into the IR interface on the back of the detector. Refer to Figure 9. a Caution Only calibrate in a fresh air environment and in a safe area. Do not calibrate the detector during or immediately after charging. Note Calibration can be aborted at any time.
GasAlertMicroClip XT Calibration 8. Ensure the sensors to be calibrated are enabled in Fleet Manager II. 9. Using , select the concentration value(s) in the Cal Gas field for each sensor. The values entered in Fleet Manager II must match the gas concentration values on the gas cylinder. 10. Click detector. to save the settings to the Calibration Procedure a Caution Only calibrate in a fresh air environment and in a safe area. Do not calibrate the detector during or immediately after charging. 1.
GasAlertMicroClip XT Technical Reference Guide Auto Zero and Oxygen Sensor Calibration Note Auto Span 4. When auto zero is complete, APPLY GAS displays. Do not apply calibration gas until APPLY GAS displays, otherwise the auto zero function will fail. 3. flashes while the detector zeroes the combustible and toxic sensors, and calibrates the oxygen sensor. 5. Attach the calibration cap to the detector. Refer to Figure 8. Open the valve on the regulator and apply gas at a flow rate of 250-500 ml/min.
GasAlertMicroClip XT Calibration When a sufficient amount of gas has been detected (approximately 30 seconds after the gas has been applied), the detector beeps once, flashes, and K remains lit while the detector completes the span (approximately 2 minutes). Successful Span If the sensors have spanned successfully, the detector beeps and the calibration procedure continues. Unsuccessful Span If any sensors fail the span, the following screen displays. Refer to Calibration Troubleshooting.
GasAlertMicroClip XT Technical Reference Guide 6. After calibration is complete, CAL DUE displays and all successfully calibrated sensors automatically reset the calibration due dates according to the calibration intervals in Fleet Manager II. Failed Sensor Past Calibration Due Date: If a sensor fails the span and it is past the calibration due date, the following three screens display. 7. The number of days that displays is when the next sensor calibration should be peformed.
GasAlertMicroClip XT Calibration Verification 1. After calibration is complete and the detector returns to normal operation, verify the calibration using a gas cylinder other than the one used for calibration. 2. The gas concentration should not exceed the sensor's detection range. Confirm the LCD shows the expected concentration. 3. To ensure the readings are accurate, apply the verification gas for the same amount of time as was applied to the sensor when it was calibrated.
GasAlertMicroClip XT Technical Reference Guide Datalogs Software Requirements The detector records various information that can be compiled to create a report. The detector is capable of storing 16 hours of information (when recording a datalog every 15 seconds). When the memory is full, the detector replaces the oldest datalogs with the most recent datalogs.
GasAlertMicroClip XT Maintenance Battery Cautions Charging the Battery a Warning To avoid personal injury and/or property damage, adhere to the following: To charge the battery, refer to Figure 10., Table 12., and the following procedures (1-8). • The detector must be deactivated to charge the battery. • Charge the battery immediately when the detector emits a low battery alarm. Refer to Charging the Battery.
GasAlertMicroClip XT Technical Reference Guide Table 12. Connecting the Charging Adapter Item Description 1 Detector 2 IR and charger interface 3 Charging adapter 4 Charging cable a Warning The detector must be charged in a safe area that is free of hazardous gas in temperatures of 32°F to 113°F (0°C to 45°C). 1. Deactivate the detector. 2. Plug the charging adapter into an AC outlet. a Caution The charging adapter is voltage specific to your region.
GasAlertMicroClip XT Maintenance Replacing a Sensor or Sensor Filter a Warning To avoid personal injury, only use sensors that are specifically designed for the detector. Refer to Replacement Parts and Accessories. Use proper ESD handling practices. • Each sensor has a high degree of resistance to common vapors and gases. To clear a sensor, move the detector to a non-hazardous environment and wait 10 to 30 minutes.
GasAlertMicroClip XT Technical Reference Guide Table 13. Replacing a Sensor or Sensor Filter Item Description 1 Front shell 2 Combustible (LEL) sensor 1. Note the placement of the PCB to ensure it is replaced correctly. Remove the two screws on the PCB. Remove the PCB carefully. a Caution Ensure no damage occurs to the battery. 3 PCB 4 PCB screws (2) 2. Remove the old sensor filter. It may be stuck to the sensors. 5 Rear shell 3. Insert the new sensor filter.
GasAlertMicroClip XT Maintenance Note Detectors that are configured for 1, 2, or 3 gases may contain a dummy sensor in one of the four sensor locations. 1. Gently remove the circular rigidified flex PCB atop the sensor from the metal sensor posts. Take care not to tear the flex cable. 3. Insert the new sensor(s). For sensor positioning, refer to Figure 12. Figure 12. Sensor Positioning 4. To complete the detector, refer to Reassembling the detector. Replacing the Oxygen Sensor 2.
GasAlertMicroClip XT Technical Reference Guide 5. Carefully replace the circular rigidified flex PCB atop the metal sensor posts. Ensure the purple plastic sensor post is inserted into the clear plastic hole. Take care not to tear the flex cable. 6. Press down to secure the circular rigidified flex PCB atop the metal sensor posts.Take care not to press down too hard and accidentally activate the detector. 7. To complete the detector, refer to Reassembling the detector. Reassembling the detector 1.
GasAlertMicroClip XT Troubleshooting Troubleshooting If a problem occurs, refer to the solutions in the Troubleshooting section. If the problem persists, contact BW Technologies by Honeywell Table 14. Troubleshooting Problem Possible Cause Solution Startup The detector does not activate. The detector enters alarm immediately when activated. The activation self-test fails. Detector automatically deactivates during startup. Depleted battery Charge battery. Refer to Charging the Battery.
GasAlertMicroClip XT Technical Reference Guide Table 14. Troubleshooting Problem Possible Cause Solution Detector Operation Detector does not display expected gas readings after activation self-test. Detector does not respond to pushbutton. Detector does not accurately measure gas. Detector does not enter into alarm. 54 Sensor not stabilized Used sensor: wait 60 seconds New sensor: wait 5 minutes Sensor(s) requires calibration Calibrate the sensor(s). Refer to Calibration.
GasAlertMicroClip XT Troubleshooting Table 14. Troubleshooting Problem Possible Cause Solution Ambient gas levels are near alarm setpoint or the sensor is exposed to a puff of the target gas Detector is operating normally. Use caution in suspect areas. Check the peak (maximum) gas exposure reading. Alarm setpoints defined incorrectly Reset alarm setpoints. Refer to Factory Gas Alarm Setpoints and Sensor Configuration. Detector requires calibration Calibrate the sensors. Refer to Calibration.
GasAlertMicroClip XT Technical Reference Guide Startup Troubleshooting Error Screen 56 Problem Solution Error Screen Problem Solution Sensor Error Calibrate the sensor(s). The sensor failed dur- Refer to Calibration. Reactivate the detector. If error ing the self-test. displays again, replace the sensor. Refer to Replacing a Sensor or Sensor Filter. IR Lock Enabled If the IR Lock screen displays, an IR device is required to calibrate the sensors.
GasAlertMicroClip XT Calibration Troubleshooting Calibration Troubleshooting Error Screen Problem Solution Auto-zero Unsuccessful H2S, CO, or LEL sensor fails to auto-zero, or O2 sensor fails to calibrate. Error Screen Problem Solution Attempt calibration again. Refer to Calibration. If ERROR displays again, replace the sensor. Refer to Replacing a Sensor or Sensor Filter. No Gas Detected If the applicable gas is not detected within 2 minutes, the detector fails the sensor.
GasAlertMicroClip XT Technical Reference Guide Replacement Parts and Accessories Model No. a Warning To order parts or accessories listed in the following table, contact BW Technologies by Honeywell. CG-Q58-4 Quad gas cylinder: CH4 (2.5%), O2 (18.0%), H2S (25 ppm), CO (100 ppm), bal. N2 (58 l) 1 CG-Q34-4 Quad gas cylinder: CH4 (2.5%), O2 (18.0%), H2S (25 ppm), CO (100 ppm), bal. N2 (34 l) 1 CG-T34 Dual gas cylinder: 50% LEL (CH4 2.5%) O2 (20.9%), bal.
GasAlertMicroClip XT Replacement Parts and Accessories Model No. Description Qty Model No. Description Qty GA-PA-3 12-24 VDC direct-wire power adapter 1 Sampling/Testing Equipment GA-PA-1* Replacement power adapter 1 GA-SPAK02 SamplerPak (motorized sampling pump kit) 1 GA-VPA-1 Vehicle charging kit 1 MC-AS01 Manual aspirator pump kit with probe (1 ft. / 0.
GasAlertMicroClip XT Technical Reference Guide Specifications Instrument dimensions: 11.25 x 6.00 x 2.89 cm (4.4 x 2.4 x 1.1 in.) Weight: 170 g (6.0 oz.) Operating temperature: -20°C to +58°C (-4°F to +136°F) +50°C to +58°C is certified by CSA International on the combustible sensor with ±5% accuracy Storage temperature: -40°C to +50°C (-40°F to +122°F) Operating humidity: 0% to 95% relative humidity (non-condensing) Alarm setpoints: May vary by region and are user defined.
GasAlertMicroClip XT Specifications Approvals: Approved by CSA to both U.S. and Canadian Standards CAN/CSA C22.2 No. 157 and C22.2 152 ANS/UL – 913 and ANSI/ISA – S12.13 Part 1 CSA Class I, Division 1, Group A, B, C, and D ATEX CE 0539 g II 1 G Ex ia IIC T4 Ga KEMA 06ATEX0056 EN 60079-0, EN 60079-11, EN 60079-26 IECEx Ex ia IIC T4 Ga IECEx CSA 05.
GasAlertMicroClip XT Technical Reference Guide 62
iERP: 131475-L3 D6568/0 [English] © BW Technologies 2010. All rights reserved.