C.B.S. Scientific 25 Adjustable Height Sequencer
5.3 Comb/Well Maximum Loading Volumes-continued
SharktoothCombs for 20cm wide units
Note: To calculate sample well volume expressed in millimeters (mm) divide maximum volume by tooth depth.
-Maximum Volume per tooth (microliters-l)
# of wells Tooth
width/mm
0.2mm thickness 0.4mm thickness 0.5mm thickness
1 158.7mm
317.4l 634.8l 793.5l
24 5.92mm
4.5l 9.0l 11.3l
34* 4.5mm
3.3l 6.6l 8.4l
38 4.02mm
2.8l 5.6l 7.0l
52* 3.0mm
2.0l 4.0l 5.0l
70* 2.25mm
1.4l 2.8l 3.4l
*Designed for multi-channel pipettor loading Point to Point Measurements:
24 well=5.92mm
34 well=4.5mm
38 well=4.02mm
52 well=3.00mm
70 well=2.25mm
Sharktooth Combs for 33cm wide units
Note: To calculate sample well volume expressed in millimeters (mm) divide maximum volume by tooth depth.
-Maximum Volume per tooth (microliters-l)
# of wells Tooth
width/mm
0.2mm thickness
0.4mm thickness
0.5mm thickness
1 284mm
539.6l 1079.2l 1349l
46 5.92mm
4.5l 9.0l 11.3l
62* 4.5mm
3.3l 6.6l 8.4l
68 4.02mm
2.8l 5.6l 7.0l
94* 3.0mm
2.0l 4.0l 5.0l
126* 2.25mm
1.71l 3.4l 4.2l
*Designed for multi-channel pipettor Point to Point Measurements:
46 well=5.92mm
62 well=4.50mm
68 well=4.02mm
94 well=3.00mm
126 well=2.25mm