Instructions Manual
IFU-20D001-20220913
Page 19
Cleaning Fundus Link
When you perform the detailed cleaning steps, please note the following:
• Ensure the cleaning wipe is not dripping, but is saturated with cleaner solution.
• Always replace the wipe when visibly dust.
Fundus Link Body:
⚫ Using a soft dry cotton cloth wipe the body gently.
Fundus Link Lens:
⚫ Disconnect the USB cable from the device.
⚫ Remove the eyecup from the device.
⚫ Using a commercially blower to blow away dust when it adheres to the lens.
⚫ And using a soft cotton cloth wipe the lens gently in a circular motion.
Fundus Link Eyecup:
⚫ Recommend to clean the eyecup before each use on a new patient
⚫ Disinfect the eyecup with a soft cloth moistened with alcohol (e.g. 70% ethanol); or
⚫ Soak the eyecup in a glutaraldehyde-based solution or hydrogen peroxide and peracetic acid solution
⚫ Rinse the eyecup under running water
⚫ Dry the eyecup (e.g. with clean paper towel) before subsequent use