OPERATION OF CRYO-CYL 120 /180 / 230 LP
24
1. Connect the transfer hose to the liquid valve
(
Item 1) of the cylinder.
2. Connect or place the other end of the hose
o
nto the inlet of the cryogenic equipment that
will receive liquid. Atmospheric dewars are
filled with a phase separator mounted to the
open end of the hose.
3. Refer to the receiving equipment manual for
procedures to open the fill valve and vent valve
of the receiving equipment.
MODEL* NITROGEN OXYGEN ARGON
Cryo-Cyl 120 LP 201 Lbs. 285 Lbs. 351 Lbs.
Cryo-Cyl 180 LP 327 Lbs. 465 Lbs. 573 Lbs.
Cryo-Cyl 230 LP 401 Lbs. 570 Lbs. 702 Lbs.
MODEL* NITROGEN OXYGEN ARGON
Cryo-Cyl 120 LP 91 Kg. 129 Kg. 159 Kg.
Cryo-Cyl 180 LP 148 Kg. 211 Kg. 260 Kg.
Cryo
-Cyl 230LP 182 Kg. 258 Kg. 318 Kg.
Note:
Filling weights are shown as the maximum weight recommended by code. Their related volumes may vary with product density.
* Relief valve setting at 22 psig (1.5 BAR)
STANDARD FILLING WEIGHT TABLE
METRIC FILLING WEIGHT TABLE
4. Open the liquid valve (Item 1) on the liquid
c
ylinder. This valve can be adjusted to obtain
the proper liquid flow rate.
5
. When the transfer is complete, close the re-
ceiving equipment's valve. Close the liquid
valve (Item 1) on the cylinder and relieve pres-
sure from the hose.
6. Disconnect or remove the hose from the re-
ceiving equipment.