User Manual 2253i / 2253iX
VOLT FREQ MAX
This test will set the voltage and frequency to a level defined by Table 9-9. The test will last for 30
VOLT UNBALANCE
minutes. The test will be repeated using the voltage from Table 9-8 and frequency from Table 9-9.
The BACK key (backspace) will terminate the test at any time.
Standard/Group RTCA A(CF) A(NF) A(WF)
Voltage offset 8 8 8 10
Frequency 400 360/440 360/650 360/800
Standard/Group Group1 Group2 Group3
Voltage offset 8 8 12
Frequency 400 360/800 360/650
Table 9-10: Emergency Voltage Unbalance
and B to 122V and phase C to a voltage lower by
a value specified by an offset. Refer to Table 9-10 for the offset value and the Frequency. The test
will repeat with the same frequency and phase A and B volt is set to 100V and phase C set to a higher
voltage specified by the offset. The test will last 30 minutes. The test will be repeated for a second
Frequency if applicable. The test can be terminated at any time.
The BACK key (backspace) will terminate the test at any time.
This test will change the output voltage for phase A