9.5.3 ABNORMAL TEST
From the DO160 MENU Scroll to the ABNORMAL AC entry using the up and down cursor keys. Pres
the ENTER key to select the ABNORMAL screens. The screen will appear as shown in Figure 9-9.
s
OLT DROP
TRANSIENTS
he up and down key
Figure 9-9: Abnormal Screen
The ABNORMAL SCREEN has the following tests:
1 VOLT MAX
2 VOLT MIN
3 VOLT UNBALNCE
4 VOLT SURG
5 V
6 FREQ
The above test can be selected by scrolling to the highlighted selection using t
and the ENTER key to start the selected test.
VOLT MAX
Standard/Group RTCA Group1/A(CF) Group2/A(NF) Group3/A(WF)
1 97 97 104/100 97 97 Voltage
3 98.5 98.5 105.5/101.5 98.5 98.5
Frequency 400 400 370 360 360
Table 9-11: Abnormal Voltage Minimum
Standard/Group RTCA Group1/ACF) Group2/A(NF) Group3/A(WF)
1 134 134 122 134 134 Voltage
3 132.5 132.5 120.5 132.5 132.5
Frequency 400 400 430 650 800
Table 9-12: Abnormal Voltage Maximum
This test will set the voltage and frequency to levels defined by Table 9-11 for 5 minutes. The test will
be repeated for Group1 and A(CF) only as indicated in Table 9-10 for voltage and Table 9-12 for
frequency. All Groups will repeat the test using Table 9-10 for the voltage setting and Table 9-10 or
Table 9-11 for the frequency setting. The BACK key (backspace) will terminate the test at any time.
California Instruments 124