Specifications

Performing a Control Solution Test
Using Control Solutions properly is important for accurate
testing.
You should test with control solutions:
®
When you first use your microdot meter. When
you can do three tests in a row correctly, you will be
able to use the meter successfully.
When you begin using a new vial of
strips.
When your test results are not
consistent with your symptoms, or
when you think your results are not
accurate.
If the meter is dropped.
Practice with control solution before you test your
®
blood with the microdot System for the first time.
When you can do three tests in a row correctly with
the control solution you are ready to test with your
own blood.
8