SUCTION UNIT NEW ASKIR 230 12V BR USER MANUAL 1 www.medicalsuction.co.
NEW ASKIR 230/12V BR Surgical Aspirator is a portable unit designed for the aspiration of bodily fluids in adults and children. It is particularly suitable for nasal, oral and tracheal aspiration of mucus, catarrh or blood after minor surgical procedures and can be used in postoperative therapy at home or conveniently transported from one hospital ward to another. The device operates on a 230V ~ AC 12V DC mixed power supply and is designed for virtually continuous use.
9. The lead battery integrated in the device is not to be considered as an ordinary domestic waste. Such a component must be disposed of in a specific collection centre in order to be recycled. 10. None of electric or mechanical parts have been designed to be repaired by customers or end-users. Don’t open the device, do not mishandle the electric / mechanical parts.
SYMBOLS Class II isolation equipment CE marking in conformity with EC directive 93/42/EEC and subsequent changes Warning, consult the instruction manual Keep in a cool, dry place Conservation temperature: -40 ÷ 70°C Type B equipment Fuse DEHP Phthalates (Suction catheter) ~ Hz I O Alternate Current Mains Frequency ON OFF Please note technical specifications may vary upon the manufacturer’s discretion! Guidance and manufacturer’s declaration – Electromagnetic Emissions The surgical aspirator NEW ASKIR 23
Guidance and manufacturer’s declaration – Electromagnetic Immunity The surgical aspiratorNEW ASKIR 230/12V BR is intended for use in the electromagnetic environment specified below. The customers or the user of the surgical aspirator NEW ASKIR 230/12V BR should assure that it’s used in such an environment. Immunity Test Compliance Electromagnetic environments guidance Electrostatic discharge (ESD) ± 6kV on contact Floors should be wood, concrete or IEC/EN 61000-4-2 ± 8kV in air ceramic tile.
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should comply with ISO 10993-1 standards on material biocompatibility. CLEANING OF ACCESSORIES To clean the plastic housing of the device wear disposable latex gloves and clean with denaturated alcohol or hypoclorite solutions.
MAINTENANCE The NEW ASKIR 230/12V BR suction equipment does not need maintenance or lubrication. It is, however to inspect the unit before each use. Unpack the instrument and always check integrity of plastic parts and feeding cable, they might have been damaged during previous use. Connect cable to electrical network and turn switch on. Close the aspirator outlet with your finger and with suction regulator in maximum vacuum position check that the vacuum indicators reaches – 75 kPa (-0.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING, PLEASE CONTACT CA-MI TECHNICAL SERVICE. THE MANUFACTURER DOES NOT GIVE GUARANTEE IF INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO BE TAMPERED. INSTRUCTIONS · Place the unit on a flat, horizontal surface. · Connect the short silicon tube, with antibacterial filter, to the suction connector .
RULES FOR RETURNING AND REPAIRING COMPLYING WITH THE NEW EUROPEAN RULES, CA-MI INDICATES THE IMPORTANT POINTS TO PROTECT INSTRUMENT AND OPERATORS HYGIENE. THESE RULES MUST BE RESPECTED IN ORDER TO GUARANTEE HYGIENE AND SAFETY TO ALL THE PEOPLE OPERATING WITH THE INSTRUMENT TO OBTAIN QUALITY AND WELL BEING. CA-MI warrants it’s products for 24 months after purchasing date.