SUCTION UNIT NEW ASKIR30 12V USER MANUAL www.medicalsuction.co.
NEW ASKIR 30 12V Surgical aspirator is a portable unit, working with 12V dc (autotration battery), designed for the aspiration of bodily fluids in adult and children. It’s particularly suitable for nasal, oral or tracheal aspiration of mucus, catarrh or blood after minor surgical procedures and can be used in post-operative therapy at home or conveniently transported from one hospital ward to another.
IMPORTANT INFORMATION FOR CORRECT DISPOSAL OF THE PRODUCT IN ACCORDANCE WITH EC DIRECTIVE 2002/96/EC: In respect of art. 13 Decreto Legislativo 25 Luglio 2005, n.151 “Actuation of European directives 2002/95/EC, 2002/96/EC and 2003/108/EC, for reduction in use of dangerous substances in the electric and electronic device and for garbage disposal” The symbol as over applied on the device or its packaging means that at the end of its useful life the product must not be disposed of with domestic waste.
Guidance and manufacturer’s declaration – Electromagnetic Emissions The surgical aspirator NEW ASKIR 30 12V is intended for use in the electromagnetic environment specified below.
ACCESSORIES SUPPLIED DESCRIPTION COMPLETE ASPIRATION JAR 1000cc CONICAL FITTING TUBES SET 6 mm x 10 mm ASPIRATION PROBE CH20 ANTIBACTERIAL FILTER Antibacterial Filter: The filter is produced with (PTFE) hydrophobic material witch prevents fluids entering the pneumatic circuit. The filter is for a single patient use which will protect patients and machines from cross contamination. When the filter is wet, it’s not possible to use the unit therefore the filter should be changed immediately.
MAINTENANCE The NEW ASKIR 30 12V suction equipment does not need maintenance or lubrication. It is, however, necessary to inspect the unit before each use. With regard to training, given the information contained in the user manual and since it is easy to understand the said device, it doesn’t appear to be necessary. Unpack the instrument and always check integrity of plastic parts and feeding cable, they might have been damaged during previous use.
WARNING: Ensure that the FLUID SIDE or IN marker on the filter is on the side facing the collection jar lid and fitted into the “VACUUM”.A wrong connection causes immediate destruction in case of contact with sucked liquids. Connect the long silicone tube to the “PATIENT” jar outlet Connect the other end of the long silicon tube to the probe plastic connector then connect the suction probe to it. Connect the power cord to the device then connect the plug to the electrical mains supply.
RULES FOR RETURNING AND REPAIRING UNDER NEW EUROPEAN RULES, CA-MI REQUIRES THE FOLLOWING PROCEDURES TO BE CARRIED OUT TO PROTECT THE INSTRUMENT AND THE SAFETY OF ALL WHO COME IN CONTACT WITH IT. CA-MI warrants it’s product for 24 months after purchasing date. Before returning an instrument for repair, the external surfaces and all accessories MUST be carefully disinfected with a cloth soaked in methylated spirits or hypochlorite-based solution.