'Define Data Tables
DataTable(Table1,True,-1)
DataInterval(0,60,Min,10)
WindVector (1,WS_ms,WindDir,FP2,False,0,0,0)
FieldNames("WS_ms_S_WVT,WindDir_D1_WVT,WindDir_SD1_WVT")
EndTable
'Main Program
BeginProg
Scan(5,Sec,1,0)
'Default Datalogger Battery Voltage measurement Batt_Volt:
Battery(Batt_Volt)
'05103 Wind Speed & Direction Sensor measurements WS_ms and WindDir:
PulseCount(WS_ms,1,1,1,1,0.098,0)
BrHalf(WindDir,1,mV2500,1,1,1,2500,True,0,_60Hz,355,0) 'mV5000
'range, 5000 mV excitation for CR3000 and CR5000 dataloggers
If WindDir>=360 Then WindDir=0
If WindDir<0 Then WindDir=0
'Call Data Tables and Store Data
CallTable(Table1)
NextScan
EndProg
7.4.4.2 CR10X Example Program
;{CR10X}
*Table 1 Program
01: 5.0000 Execution Interval (seconds)
1: Pulse (P3)
1: 1 Reps
2: 1 Pulse Channel 1
3: 21 Low Level AC, Output Hz
4: 3 Loc [ WS_ms ]
5: 0.098 Multiplier
6: 0 Offset
2: Excite-Delay (SE) (P4)
1: 1 Reps
2: 5 2500 mV Slow Range ; 5000 mV(slow/60 hz) Range for CR23X, 21X, CR7
3: 1 SE Channel
4: 1 Excite all reps w/Exchan 1
5: 2 Delay (0.01 sec units)
6: 2500 mV Excitation ; 5000 mV for CR23X, 21X, CR7
7: 4 Loc [ WindDir ]
8: 0.142 Multiplier ; 0.071 for CR23X, 21X, CR7
9: 0 Offset
3: If (X<=>F) (P89)
1: 4 X Loc [ WindDir ]
2: 3 >=
3: 360 F
4: 30 Then Do
4: Z=F x 10^n (P30)
1: 0 F
2: 0 n, Exponent of 10
3: 4 Z Loc [ WindDir ]
13