Model 092 Barometric Pressure Sensor Revision: 8/12 C o p y r i g h t © 2 0 1 2 C a m p b e l l S c i e n t i f i c , I n c .
Warranty “PRODUCTS MANUFACTURED BY CAMPBELL SCIENTIFIC, INC. are warranted by Campbell Scientific, Inc. (“Campbell”) to be free from defects in materials and workmanship under normal use and service for twelve (12) months from date of shipment unless otherwise specified in the corresponding Campbell pricelist or product manual. Products not manufactured, but that are re-sold by Campbell, are warranted only to the limits extended by the original manufacturer.
Assistance Products may not be returned without prior authorization. The following contact information is for US and international customers residing in countries served by Campbell Scientific, Inc. directly. Affiliate companies handle repairs for customers within their territories. Please visit www.campbellsci.com to determine which Campbell Scientific company serves your country. To obtain a Returned Materials Authorization (RMA), contact CAMPBELL SCIENTIFIC, INC., phone (435) 227-9000.
Table of Contents PDF viewers: These page numbers refer to the printed version of this document. Use the PDF reader bookmarks tab for links to specific sections. 1. Introduction .................................................................1 2. Cautionary Statements...............................................1 3. Initial Inspection .........................................................1 4. Quickstart ....................................................................1 4.1 4.2 Installation.......
Table of Contents Tables 7-1. A-1. A-2. Connecting the 092 to Campbell Scientific Dataloggers..................... 8 Dip Switches SW1: Analog Output Voltage Settings ..................... A-1 Dip Switches SW2: Analog Output Pressure Range Settings .........
Model 092 Barometric Pressure Sensor 1. Introduction The Model 092 Barometric Pressure Sensor is designed to measure ambient atmospheric pressure and is configured at Campbell Scientific for analog output. It is recommended for wind-farm power-performance research. 2. 3. 4. Cautionary Statements • Install the sensor such that it faces north in the northern hemisphere, or south in the southern hemisphere. This orientation will shield the polycarbonate sensor housing from most direct sunlight.
Model 092 Barometric Pressure Sensor FIGURE 4-1. 092 installed on a vertical mast Individual wire leads connect to the CR1000 datalogger as shown in step 5 of the procedure in Section 4.2, Programming with Short Cut. A generalized wiring diagram is provided in Section 7.2, Wiring.
Model 092 Barometric Pressure Sensor 4.2 Programming with Short Cut Short Cut Program Generator for Windows can be used to program the CR1000 datalogger to measure the 092 as outlined in the following procedure. Short Cut can also program other 092-compatible Campbell Scientific dataloggers. 1. Open Short Cut and click on New Program. 2. Select a scan interval.
Model 092 Barometric Pressure Sensor 4 3. Select 092 Barometric Pressure Sensor and click the right arrow in the middle of the page to add it to the list of sensors to be measured. 4. Define the name of the public variable, or use the default BP_mmHg to hold the barometric pressure measurement. Select the desired units of measure, or use the default mmHg.
Model 092 Barometric Pressure Sensor 5. Sensor connections to the CR1000 datalogger are shown in the Wiring tab. 6. Select the desired output data for final storage and click Finish. 7. Click Wiring Diagram for a description of sensor wiring for this example. Send the program from the PC to the CR1000 datalogger when prompted if a telecommunications link is active.
Model 092 Barometric Pressure Sensor 5. Overview The 092 ships from Campbell Scientific pre-wired for analog output with the following settings: • Analog Output Voltage: 0 to 5 Vdc • Range: 600 to 1100 mbar Refer to Appendix A, Analog Output Configurations, for available analog output options. The 092 uses an on-board microcontroller and piezoresistive pressure sensor module.
Model 092 Barometric Pressure Sensor 6. Specifications Features: • • The following 092 barometer settings are configurable: o measurement range o data output option / communication option o analog output range Specifications listed here are limited to those applicable to the 092 as it is configured and shipped by Campbell Scientific, Inc. Compatibility: Measurement Range: Resolution: Accuracy: Long-term stability: 7. CR800 series CR1000 CR3000 CR23X CR10(X) 600 to 1100 mbar 0.1 mbar (.
Model 092 Barometric Pressure Sensor southern hemisphere. This orientation will shield the polycarbonate sensor housing from most direct sunlight. 7.2 Wiring TABLE 7-1. Connecting the 092 to Campbell Scientific Dataloggers 7.
Model 092 Barometric Pressure Sensor 'Main Program BeginProg 'Main Scan Scan(5,Sec,1,0) 'Default Datalogger Battery Voltage measurement 'BattV' Battery(BattV) 'Default Wiring Panel Temperature measurement 'PTemp_C' PanelTemp(PTemp_C,_60Hz) '092 Barometric Pressure Sensor measurement 'BP_mmHg' VoltSE(BP_mmHg,1,mV5000,1,1,0,_60Hz,0.1,600) BP_mmHg=BP_mmHg*0.75006 'Call Data Tables and Store Data CallTable(Table1) CallTable(Table2) NextScan EndProg 8. Maintenance 8.
Model 092 Barometric Pressure Sensor 10
Appendix A. Analog Output Configurations The 092 ships from Campbell Scientific pre-wired and pre-configured for analog output with the following settings: • Analog Output Voltage: 0 to 5 Vdc • Range: 600 to 1100 mbar Output voltage and measurement ranges are set by switches. Dip switch settings are listed in TABLE A-1 and TABLE A-2. Highlighted settings are those set at the factory by Campbell Scientific.
Appendix A. Analog Output Configurations TABLE A-2. Dip Switches SW2: Analog Output Pressure Range Settings RANGE SELECTION mbar SWITCH SETTINGS mm Hg in. Hg LOWER UPPER LOWER UPPER LOWER UPPER SW2-1 SW2-2 SW2-3 SW2-4 600 800 450 600 17.72 23.62 ON ON ON ON 600 900 450 675 17.72 26.58 OFF ON ON ON 600 1000 450 750 17.72 29.53 ON OFF ON ON 6001 11001 4501 8251 17.721 32.481 OFF1 OFF1 ON1 ON1 700 800 525 600 20.67 23.
Appendix B. Wiring The 092 ships from Campbell Scientific pre-wired for analog output. Internal cable connections are shown in FIGURE B-1. FIGURE B-1.
Appendix B.
Appendix C.
Appendix C.
Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 North Logan, Utah 84321 UNITED STATES www.campbellsci.com • info@campbellsci.com Campbell Scientific Africa Pty. Ltd. (CSAf) PO Box 2450 Somerset West 7129 SOUTH AFRICA www.csafrica.co.za • cleroux@csafrica.co.za Campbell Scientific Australia Pty. Ltd. (CSA) PO Box 8108 Garbutt Post Shop QLD 4814 AUSTRALIA www.campbellsci.com.au • info@campbellsci.com.au Campbell Scientific do Brazil Ltda.