Manual

229 Heat Dissipation Matric Water Potential Sensor
3. Measurements of sensor temperature response are made periodically to
determine if equilibration is attained. This will require depressurization of
the pressure vessel if a pressure-tight feedthrough is not used. Prior to
depressurization, it is important that the effluent hose be blocked by
clamping or other method to prevent solution from re-entering the soil and
sensors.
4. When equilibration is attained, the effluent hose is blocked, the vessel is
depressurized and 229 measurements are recorded.
5. The next calibration pressure is then applied and the process repeated.
Pressure-tight bulkhead connectors are available for some pressure vessels.
Determining whether equilibrium has been reached is simplified when using a
feedthrough connector since the pressure vessel doesn’t have to be
depressurized and opened for each reading. A temporary connector can be
used to disconnect the datalogger from the other components installed in the
pressure vessel.
Pressurized readings can be used for determining equilibrium but should not be
used for calibration data. Thermal properties are affected by pressure and
calibration data should be collected at the same pressure the sensors will be
used—in most cases this is atmospheric pressure.
6.4.1 Wiring for Calibration using Pressure Plate Extractor
The wiring arrangement of Figure 6-4 depicts a datalogger with an AM16/32
multiplexer, a CE4 current source and a thermistor being used in a typical 229
calibration arrangement. This is similar to the wiring for program example #2
and #4 (see Section 5). All components except the datalogger can be placed
inside a 5 bar pressure vessel. If this method is used, the electronic
components in the pressure vessel should be protected against moisture
damage. Place the components in a container such as plastic bag with
desiccant.
25