User guide
CS210 Enclosure Humidity Sensor 
'Main Program 
BeginProg 
 Scan (10,Sec) 
 Battery (Batt_volt) 
    'Power the CS210 using C2 and leave it on. 
 PortSet (C2,1 ) 
 VoltSe (Enc_RH,1,1,0.1,0) 
    'Set the sensor to 100% if it exceeds 100%. 
    If Enc_RH > 100 Then Enc_RH = 100 
 CallTable Daily 
 NextScan 
EndProg 
5.2 Edlog Examples 
Example 3 is a portion of a CR10(X) program, but our CR500, CR510, and 
CR23X dataloggers are programmed similarly. Wiring used for the CR10(X) 
program is included in Table 3. 
Example 4 is an Edlog program written for a 21X Micrologger; wiring used for 
the 21X program is included in Table 3. The 21X program turns on control 
port 1 to power the CS210. Notice that the scan rate is set at 10 seconds. The 
sensor won’t give a good reading until the second scan. 
TABLE 3. Wiring for Examples 3 and 4 
Description  Color  CR10(X)  21X 
Relative Humidity  White  SE 6 (3L)  SE 1 (1H) 
Signal & Power Reference  Clear  G  G 
Power Black 5 VDC C1 
Example 3. Sample CR10(X) Program 
;Measure the CS210 relative humidity. 
; 
01: Volt (SE) (P1) 
 1: 1   Reps 
  2:  5    2500 mV Slow Range  ;CR510 (2500 mV); CR23X (1000 mV) 
 3: 6   SE Channel  ;White wire (SE 6), Clear wire (G) 
  4:  3    Loc [ RH_enc ] 
 5: .1   Mult 
 6: 0   Offset 
5 










