Appendix B. SDI-12 Commands/Changing Settings 
FIGURE B-4. SDI-12 transparent mode on CR10X datalogger using 
control port 1 and changing SDI-12 address from 0 to 1 
B.2.5 CR10X-PB Table-Based Datalogger Example 
1.  Connect a single sensor to the datalogger as follows: 
•  White to Control Port C1 
•  Black, Orange, Clear to G 
•  Red to 12V 
2.  Download a datalogger program that contains the SDI-12 Recorder (P105) 
instruction with valid entries for each parameter. Make sure that 
parameter 3 of the P105 instruction matches the control port number where 
the green wire is connected. 
3.  In the LoggerNet Connect screen navigate to the Datalogger menu and 
select Terminal Emulator. The “Terminal Emulator” window will open. 
In the Select Device menu, located in the lower left-hand side of the 
window, select the CR10XTD or CR10XPB station. 
4.  Click on the Open Terminal button. 
5.  Press the <enter> key until the datalogger responds with the “>” prompt. 
6.  To activate the SDI-12 Transparent Mode on Control Port p, enter *8. The 
TD datalogger will respond with a “.” prompt. At the “.” prompt enter #. 
The TD datalogger will respond with 150000. Finally, enter p (Control 
Port p) and press the <enter> key. For this example, p = 1. The TD 
datalogger will respond with “entering SDI-12”. If any invalid 
SDI-12 command is issued, the datalogger will exit the SDI-12 
Transparent Mode. 
7.  To query the sensor for its current SDI-12 address, enter the command ?!. 
The sensor will respond with the current SDI-12 address. 