User Manual

NR-LITE Net Radiometer
2: If time is (P92)
1: 0 Minutes (Seconds --) into a
2: 60 Interval (same units as above)
3: 10 Set Output Flag High (Flag 0)
3: Real Time (P77)
1: 110 Day,Hour/Minute (midnight = 0000)
4: Aver age (P71 )
1: 1 Reps
2: 1 Loc [ Net_rad ]
5.3.4 CR10X Example Program with Wind Speed Correction
This example measures the NR-LITE using a single-ended input and
Instruction 1. The pro gram also measures wind speed and appl i e s the
correction factor as described in Section 1.4. Average net radiation and wind
speed are output every hour.
NR-LITE Wiring for Example 4
Color
Function
Example CR10X
Program Channels
Used
White Radiation Signal S.E. Channel 1
Green Signal Reference AG
Clear Shield G
;{CR10X}
;
*Table 1 Program
01: 60 Execution Interval (seconds)
;measure 05103 wind speed
1: Pulse (P3)
1: 1 Reps
2: 1 Pulse Channel 1
3: 21 Low Level AC, Output Hz
4: 1 Loc [ Wspd_m_s ]
5: .0980 Mult
6: 0 Offset
;measure NR-LITE net radiation
2: Volt (SE) (P1)
1: 1 Reps
2: 3** 25 mV Slow Range
3: 1* SE Channel
4: 2* Loc [ Rn_obs ]
5: 63.29 Mult ;Multiplier for specific sensor
6: 0 Offset ;in units of (W m
-2
)/mV (see above)
13