Instruction Manual
NR01 Four-Component Net Radiation Sensor 
The PT-100 sensor can connect directly to the CR3000 and CR5000 
dataloggers because they have current excitation outputs. Refer to Table 4.3-4 
and Program Example 5.2.2 for information on using the current excitation 
technique with a CR3000 or CR5000 datalogger. 
TABLE 4.3-4. CR3000 and CR5000 Connections 
for Differential Measurement and using the Current Excitation to 
Measure the PT100 Sensor 
Wire Label  Color  CR3000/CR5000 
Pyranometer Up Sig 
Red 
Differential Input (H) 
Pyranometer Up Ref 
*Blue 
Differential Input (L) 
Pyranometer Down Sig 
White 
Differential Input (H) 
Pyranometer Down Ref 
*Green 
Differential Input (L) 
Pyrgeometer Up Sig 
Brown 
Differential Input (H) 
Pyrgeometer Up Ref 
*Yellow 
Differential Input (L) 
Pyrgeometer Down Sig 
Purple 
Differential Input (H) 
Pyrgeometer Down Ref 
*Grey 
Differential Input (L) 
PT100 Signal 
**White 
Differential Input (H) 
PT100 Signal Ref 
**Green 
Differential Input (L) 
Current Excite  **Red  Current Excitation IX 
Current Return -  **Blue  Current Excitation IXR 
Shield (both cables)  Clear 
*Jumper to AG or   with user-supplied wire. 
**Note these are in Cable 2. 
4.3.1 Connecting and Using the Heater 
Only use the sensor heater when there is risk of dew forming on the sensors, 
especially for low power installations. Furthermore, the heater should be 
turned on and off infrequently as it may take some time for the sensor to come 
to thermal equilibrium. No damage will result if the heater is powered 
permanently, but as with all thermopile sensors, it is best if the sensor operates 
at ambient temperatures and is not subject to rapidly changes of temperature. 
The sensor power can be controlled using one of the 12V power switches built 
into Campbell dataloggers or using an external solid-state switch such as a 
PSW12/SW12. The heater current drain is approximately 140 mA from a 12 V 
battery. Connect the ground return from the heater, either directly to the 
battery, or to a G terminal close the power input to the logger (i.e., not to an 
analog ground near the measurement inputs). 
The heater power can be controlled by adding instructions to the datalogger 
program, that turns on the heater only when the light level falls below 20 W m
-
2
 or, if a measurement of air humidity is available, when the dew point of the 
air falls to within 1ÂșC of the sensor body temperature. Appendix A provides an 
example CR3000/CR5000 program that controls the NR01 heater. 
19 










