Instruction manual
9522B Iridium Satellite Modem 
_____________________________________________________________________________ 
19 
Figure 19: Generic - Modem Configuration 
3.1.7.  Step 7: PakBusPort Configuration 
As in Figure 20, be sure the Communications Enabled box is 
checked, change the Maximum Baud Rate to 19200 (9600 for a 
CR10X), and add a 200 ms delay under Delay Hangup. You may 
change other settings in this configuration to suit your particular 
application. Once complete move to Step 8 in order to configure 
the datalogger. 
Figure 20: PakBus Port Configuration 










