Specifications
AM25T Solid State Multiplexer 
9 
Input locations within the measurement loops must be indexed (--), see 
Instruction 87 in the datalogger manual. To index a location, using Edlog ver 
6.0 or greater, move the cursor to the Input location label and press “F4”. To 
index a location with the datalogger keyboard, press the “C” key when 
entering the location number and before the “A” key is pressed. 
TABLE 6-1. RTD Excitation Voltage and Measurement Range for AM25T 
Multiplexers Prior to Serial Number 1839 
Datalogger  Temperature Range  Excitation voltage  Vs Range  Vx Range 
21X  -40 to 85°C 
-24 to 65°C 
250 mV 
350 mV 
± 5 mV 
± 5 mV 
± 500 mV 
± 500 mV 
CR10(X)  -40 to 85°C 
-11 to 52°C 
350 mV 
250 mV 
± 7.5 mV 
± 2.5 mV 
± 2500 mV 
± 250 mV 
CR7  -40 to 85°C  350 mV 
± 15 mV  ± 500 mV 
TABLE 6-2. Wiring for CR10(X) Examples 
Function AM25T CR10(X) 
+12V Power  12 V  12 V 
Power and Shield Ground 
G 
Clock CLK C1 
Reset RES C2 
RTD Excitation  EX  E1 
Analog Ground  AG  AG 
Common High  HI  1H 
Common Low  LO  1L 
Example 1. CR10(X) Program for Measuring 25 Type T Thermocouples or Voltage Sensors 
Using a Differential Instruction 
*Table 1 Program 
 01: 10   Execution Interval (seconds) 
01: Set Port(s) (P20)  ;Configure Control Ports for 1 millisecond pulse 
 1: 9999   C8..C5 = nc/nc/nc/nc 
 2: 9933   C4..C1 = nc/nc/1ms/1ms 
02: Do (P86)      ;Turn On AM25T 
  1:  42    Set Port 2 High 
03: Full Bridge (P6)  ;Measure the output of the reference temp. full bridge 
 1: 1   Reps 
  2:  23    25 mV 60 Hz Rejection Range  ;See Table 6-1 for older AM25T multiplexers 
 3: 1   DIFF Channel 
  4:  1    Excite all reps w/Exchan 1 
 5: 1200   mV Excitation  ;See Table 6-1 for older AM25T multiplexers 
  6:  1    Loc [ RefTemp_C ] 
 7: -0.001   Mult 
 8: 0.09707   Offset 