Specifications
AM25T Solid State Multiplexer 
19 
ClkPort: The ClkPort argument is the control port that will be used to clock 
the AM25T. One clock port may be used with several AM25Ts. A numeric 
code is entered for this argument: 
Code
 Description 
1  Control Port 1 
2  Control Port 2 
3  Control Port 3 
4  Control Port 4 
5  Control Port 5 
6  Control Port 6 
7  Control Port 7 
8  Control Port 8 
ResPort: The ResPort argument is the control port that will be used to enable 
and reset the AM25T. Each AM25T must have a unique Reset port. A numeric 
code is entered for this argument; refer to ClkPort above. 
ExChan: The ExChan argument is the excitation channel that will be used to 
provide switched excitation for the PRT reference temperature measurement. 
An alphanumeric or numeric code can be entered: 
Alphanumeric
 Numeric Description 
0  0  Temperature not measured 
VX1  1  Excitation channel 1 
VX2  2  Excitation channel 2 
VX3  3  Excitation channel 3 
RevDiff: A constant is entered for the RevDiff parameter to determine whether 
the inputs are reversed and a second measurement made. This function will 
remove any voltage offset errors due to the datalogger measurement circuitry, 
including common mode errors. Enabling this parameter will double 
measurement time. False (or 0) = Do not make second measurement; True (or 
1) = Reverse inputs and make second measurement. 
SettlingTime: The SettlingTime parameter is the amount of time to delay after 
setting up a measurement and before making the measurement. Refer to the 
table below for default SettlingTimes. 
Entry
 Range Integration Settling Time 
0 All 250 µs 450 µs (default) 
0  All  _50Hz  3 ms (default) 
0  All  _60Hz  3 ms (default) 
>
100 All  All  µs entered 
Integ: The Integ parameter is the amount of time, in microseconds, to integrate 
a signal for the channel being measured. 
Option Description 
250  Performs a 250 microsecond integration. 
_60Hz  Performs a 16.667 millisecond integration; filters 60 Hz noise. 
_50Hz  Performs a 20 millisecond integration; filters 50 Hz noise. 