Specifications
Appendix A. Single-Ended Measurements 
A-2 
A.1.1 Thermocouple Measurements 
Wire the high side of each thermocouple into an input terminal. The low side 
of each thermocouple is wired into the adjacent ground terminal (Figure A-1). 
FIGURE A-1. Single-ended Measurement of a Type T Thermocouple 
Example A-1. CR10 Program for Measuring 50 Type T Thermocouples or Voltage Sensors 
Using a Single-Ended Instruction 
*Table 1 Program 
 01: 10   Execution Interval (seconds) 
01: Set Port(s) (P20)  ;Configure Control Ports for 1 millisecond pulse 
 1: 9999   C8..C5 = nc/nc/nc/nc 
 2: 9933   C4..C1 = nc/nc/1ms/1ms 
02: Do (P86)      ;Turn On AM25T 
  1:  42    Set Port 2 High 
03: Full Bridge (P6)  ;Measure the output of the reference temp. full bridge 
 1: 1   Reps 
  2:  24    250 mV 60 Hz Rejection Range  ;See Table 6-1 for older AM25T multiplexers 
 3: 1   DIFF Channel 
  4:  1    Excite all reps w/Exchan 1 
 5: 1200   mV Excitation  ;See Table 6-1 for older AM25T multiplexers 
  6:  1    Loc [ RefTemp_C ] 
 7: -0.001   Mult 
 8: 0.09707   Offset 
04: BR Transform Rf[X/(1-X)] (P59)  ;Calculate RTD resistance R/R0 
 1: 1   Reps 
  2:  1    Loc [ RefTemp_C ] 
 3: 10.025   Multiplier (Rf) 