Specifications
Appendix A. Single-Ended Measurements 
A-3 
05: Temperature RTD (P16)  ;Calculate reference temperature 
 1: 1   Reps 
  2:  1    R/R0 Loc [ RefTemp_C ] 
  3:  1    Loc [ RefTemp_C ] 
 4: 1.0   Mult 
 5: 0.0   Offset 
06: Beginning of Loop (P87)  ;Loop through thermocouple channels 
 1: 0   Delay 
 2: 25   Loop Count 
07: Step Loop Index (P90)  ;Set index to advance 2 locations/loop 
 1: 2   Step 
08: Do (P86)      ;Clock the AM25T twice 
 1: 71   Pulse Port 1 
09: Do (P86) 
 1: 71   Pulse Port 1 
10: Thermocouple Temp (SE) (P13)  ;Measure the connected thermocouple 
 1: 2   Reps 
  2:  21    2.5 mV 60 Hz Rejection Range 
 3: 1   SE Channel 
 4: 1   Type T (Copper-Constantan) 
  5:  1    Ref Temp (Deg. C) Loc [ RefTemp_C ] 
  6:  2--    Loc [ TC_1 ] 
 7: 1.0   Mult 
 8: 0.0   Offset 
11: End (P95)      ;End loop 
12: Do (P86)      ;Turn off AM25T 
  1:  52    Set Port 2 Low 
Example A-2. 21X Program for Measuring 50 Type T Thermocouples or Voltage Sensors 
Using a Single-ended Instruction 
*Table 1 Program 
 01: 10   Execution Interval (seconds) 
01: Do (P86)      ;Turn On AM25T 
  1:  42    Set Port 2 High 
02: Full Bridge (P6)  ;Measure the output of the reference temp. full bridge 
 1: 1   Reps 
  2:  3    50 mV Slow Range  ;See Table 6-1 for older AM25T multiplexers 
 3: 1   DIFF Channel 
  4:  1    Excite all reps w/Exchan 1 
 5: 2400   mV Excitation  ;See Table 6-1 for older AM25T multiplexers 
  6:  1    Loc [ RefTemp_C ] 
 7: -0.001   Mult 
 8: 0.09707   Offset 