Specifications
Appendix A. Single-Ended Measurements 
A-7 
Example A-4. Measuring Single Ended Thermocouples with the Input Location Indexed and 
no Step Loop Index 
01: Beginning of Loop (P87) 
 1: 0  Delay 
 2: 6Loop Count 
02: Thermocouple Temp (SE) (P13) 
 1: 2   Reps 
  2:  21    2.5 mV 60 Hz Rejection Range 
 3: 1   SE Channel 
 4: 1   Type T (Copper-Constantan) 
  5:  1    Ref Temp (Deg. C) Loc [ RefTemp ] 
  6:  18--    Loc [ TC_1 ] 
 7: 1.0   Mult 
 8: 0.0   Offset 
03: End (P95) 
Example A-5. Measuring Single Ended Thermocouples with the Input Location Indexed and 
Step Loop Index 
01: Beginning of Loop (P87) 
 1: 0   Delay 
 2: 6   Loop Count 
02: Step Loop Index (P90) 
 1: 2   Step 
03: Thermocouple Temp (SE) (P13) 
 1: 2   Reps 
  2:  21    2.5 mV 60 Hz Rejection Range 
 3: 1   SE Channel 
 4: 1   Type T (Copper-Constantan) 
  5:  1    Ref Temp (Deg. C) Loc [ RefTemp ] 
  6:  18--    Loc [ TC_1 ] 
 7: 1.0   Mult 
 8: 0.0   Offset 
04: End (P95) 