Specifications
1 
AM25T Solid State Multiplexer 
1. General 
The AM25T Multiplexer increases the number of channels for measuring 
thermocouples or voltage sensors with Campbell Scientific dataloggers. The 
AM25T is positioned between the sensors and the datalogger. The datalogger 
controls the AM25T's solid state relays, sequentially connecting each sensor to 
the datalogger. 
The AM25T is not suitable for resistive bridge measurements (high impedance 
of the solid state relays) or multiplexing power (25 mA maximum switching 
current). 
A maximum of 25 thermocouples (differential voltage measurement) can be 
multiplexed by an AM25T. Single-ended voltage measurements are not 
recommended (Appendix A). 
The AM25T must be used in a non-condensing environment. An enclosure is 
required for field use. In applications where one or two multiplexers are 
deployed in the field, the ENC 12/14 can be used. The ENC 16/18 can be used 
to house several multiplexers at the same site. 
1.1 AM25T Specifications 
POWER 
Unregulated 9.6 V to 16 V; 12 VDC Nominal 
CURRENT DRAIN 
Quiescent: 0.5 mA 
Active:  1.0 mA (typical) 
ENABLE 
Inactive:  < 0.9 V 
Active:  3.5 to 5.0 V (7 V max.) 
CLOCK 
The relays are advanced on the falling edge of the clock pulse (transition from 
>3.5 V to <1.5 V; 7 V max.). The minimum clock pulse width is limited by 
the datalogger. 
Minimum ON time  50 microseconds 
Minimum OFF time  60 microseconds 